:GENerator#:VTERm
Syntax:
:GENerator#:VTERm <vterm>
Defines the termination voltage of the connected load. This setting only has an effect if the
termination scheme is set to
SINGle
.
This setting is required to calculate the correct output levels.
Example command:
GEN0:VTER 0.0
:GENerator#:VTERm?
Syntax:
:GENerator#:VTERm?
Returns the currently configured termination voltage.
Example query:
GEN0:VTER?
Example response:
0
:GENerator#:MODE
Syntax:
:GENerator#:MODE <DATapattern|DIVidedclock>
Defines whether the generator output transmits data from the sequencer (
DATapattern
) or a
divided clock signal (
DIVidedclock
).
Example command:
GEN0:MODE DAT
:GENerator#:MODE?
Syntax:
:GENerator#:MODE?
Returns the currently configured generator mode.
Example query:
GEN0:MODE?
Example response:
DATapattern
:GENerator#:CHANnel
Syntax:
:GENerator#:CHANnel <channel>
Defines the sequencer channel from which the generator gets its data pattern (only valid if the
generator mode is set to
DATapattern
).
This channel number corresponds to the channel that was specified when downloading a pattern
with the
:SEQ:PATT:DOWN
command.
Example command:
GEN0:CHAN 2
:GENerator#:CHANnel?
Syntax:
:GENerator#:CHANnel?
Returns the currently configured sequencer channel number.
Example query:
GEN0:CHAN?
Example response:
2
46