:GENerator#:ENABle
Syntax:
:GENerator#:ENABle <bool>
Enables or disables the generator output relay.
Example command:
GEN0:ENAB 1
:GENerator#:ENABle?
Syntax:
:GENerator#:ENABle?
Returns the current state of the output relay.
Example query:
GEN0:ENAB?
Example response:
1
:GENerator#:ERRor?
Syntax:
:GENerator#:ERRor?
Queries the output protection state of the generator output. If an error was detected, this query
returns
1
. To clear the error, the output relay must be turned on or off using the
:GEN#:ENAB
command.
Example query:
GEN0:ERR?
Example response:
0
:GENerator#:TERMination
Syntax:
:GENerator#:TERMination <OPEN|SINGle|DIFFerential>
Specifies the termination scheme of the load that is connected to the generator output:
•
OPEN
: each generator output connects to a high-impedance load
•
SINGle
: each generator output connects to a load that is terminated with 50
Ω into ground,
or into a termination voltage
•
DIFFerential
: the generator outputs are connected to a differential 100 Ω load, with no
connection to ground
This setting is required to calculate the correct output levels.
Example command:
GEN0:TERM SING
:GENerator#:TERMination?
Syntax:
:GENerator#:TERMination?
Returns the currently configured terminations scheme.
Example query:
GEN0:TERM?
Example response:
SINGle
BIT-3000 User Manual