background image

 

 

   

BT-710 Hand held pulse oximeter

 

Operation Manual

 

  

 

 

 

 

Keep this manual for future reference 

  

P/N : 710-ENG-OPM-EUR-R02

 

 

Содержание BT-710

Страница 1: ...BT 710 Hand held pulse oximeter Operation Manual Keep this manual for future reference P N 710 ENG OPM EUR R02...

Страница 2: ...er no responsibility is assumed by Bistos for its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implication or otherwise un...

Страница 3: ...Product outlook 11 1 5 Description of pulse oximeter 11 1 6 Understanding the display 12 1 7 Essential performance 13 2 Preparing for operation 13 2 1 Installation 13 2 2 Connecting to power 15 3 Basi...

Страница 4: ...ning 28 9 3 Disinfection 29 10 Maintenance 30 10 1 Checking 30 10 2 Trouble shootings 30 10 3 Maintenance plan 31 11 Accessories 31 12 Specifications 32 12 1 Safety specification 32 12 2 Hardware spec...

Страница 5: ...ghly before using BT 710 Used to identify safety information Be well known this information thoroughly before using BT 710 IPX2 Indicates the protection level against the ingress of liquid IPX2 is pro...

Страница 6: ...safety This should include leakage current measurement and insulation testing The recommended testing inter val is once per year If the hospital or healthcare institutions using this device fail to i...

Страница 7: ...vices the user a doctor or a nurse should ensure the safety of the patient and instru ment The physiological wave physiological parameters and alarm information dis played on the pulse oximeter are on...

Страница 8: ...ety If there is obvious damage or aging replace the parts before use The replacement should be made with same parts of original parts Before powering on the device make sure that the power used by the...

Страница 9: ...excessive moisture that would allow for liquid pooling Personal injury or equipment damage could occur Do not touch the patient and signal input output parts simultaneously Due to the risk of electri...

Страница 10: ...oid exposure to direct sunlight Avoid placing in an area where high variation of temperature exists Operating temperature ranges from 5 40 Operating humidity ranges from 30 85 Avoid in the vicinity of...

Страница 11: ...e oximeter is suitable for bedside monitoring of patient This pulse oximeter enables the monitoring of blood oxygen saturation SpO2 and pulse rate PR It is equipped with a replaceable built in battery...

Страница 12: ...ch screen and built in lithium ion battery Adult SpO2 sensor probe AC DC adapter 1 4 Product outlook Figure1 3 Top view Figure1 1 Front view Figure1 2 Rear view Figure1 4 Bottom view 1 5 Description o...

Страница 13: ...econds 4 Alarm reset To reset the alarm condition 5 Setting Enter to the setting mode Press again to close the setting mode Figure1 6 Top view Name Description 6 SpO2 SpO2 sensor probe interface Figur...

Страница 14: ...sion index by placing the sensors to the appropriate site of patient The device is composed with display control circuit and panel and input part for SpO2 sensors It detects SpO2 and PR using specific...

Страница 15: ...packing box and materials for use in future transporting or storage 2 1 2 Power requirements DC power supply adapter Model UE10WCP1 050200SPA Input A C 100 V 240 V 50 60 Hz Output D C 5 V 2 0 A Built...

Страница 16: ...Basic operations 3 1 Turn on 3 1 1 Check the pulse oximeter Before turn on the pulse oximeter check whether there is mechanical damage to the pulse oximeter and whether the external cables and accesso...

Страница 17: ...onds countdown window and the pulse oximeter turns off in 3 seconds CAUTION If the pulse oximeter is not turned off properly you can simply disconnect the power to shutdown forcibly But the forced shu...

Страница 18: ...r is Italian Deutsch The interface language of the monitor is German Select Default clears all the setting except the date and time and go back to the initial factory setup Select key to display the f...

Страница 19: ...sensor or sensor packaging has signs of damage do not use this SpO2 sensor return it to the manufacturer If there is carboxyhemoglobin methemoglobin or dye diluted chemical the SpO2 value will have d...

Страница 20: ...urn on the pulse oximeter plug in SpO2 probe and connect pa tient s finger or toe monitor displays SpO2 wave SpO2 Pulse Search dis played in the technical alarm area until the monitor measured SpO2 va...

Страница 21: ...per limit Select SpO2 High Limit and set the SpO2 alarm upper limit value Range 1 100 Adult Pediatric Neonate default 100 100 95 The input cannot enter less than the SpO2 alarm low limit Select PR Low...

Страница 22: ...ondensing Atmospheric pressure 700hPa 1060hPa 5 6 Technical description The SpO2 sensor probe specified in Chapter 11 material which contacts patients or other staff has passed the biocompatibility te...

Страница 23: ...rompts the medical staff through sound and light when the abnormal changes in vital signs are monitored or the pulse oximeter has a failure or is unable to monitor the patient successfully For better...

Страница 24: ...ation associated with system status in addition to the physiological alarms and technical alarms and generally such information do not involve the patient s vital signs The prompt messages generally a...

Страница 25: ...gh priority alarm Beep beep beep beep beep beep beep beep beep beep Medium priority alarm Beep beep beep Low priority alarm Beep 7 3 3 Alarm information Alarm information displayed on the physiologica...

Страница 26: ...ed Note that in this chapter P column indicates the default alarm priority H indicates high priority M indicates middle priority L indicates low priority and indicates configurable priority by the use...

Страница 27: ...the sensor mounting position whether the sensor is damaged or sensor type Reconnect the sensor or use new sensor SpO2 Sensor Off L SpO2 sensor falls off from the patient or monitor malfunctions or sen...

Страница 28: ...read this manual and labels on the battery surface carefully Do not drop the battery If it won t be used for a long time over three months please store the battery properly Charge the battery to 50 a...

Страница 29: ...e model with established battery 8 4 Battery recycling If the battery has visible damage or cannot store power it should be replaced and recycled properly Follow the appropriate regulations to dispose...

Страница 30: ...e with non etching cleaner such as soap and water CAUTION To avoid damaging the pulse oximeter Do not use strong solvents such as acetone Most cleaners must be diluted before use Diluting should be ac...

Страница 31: ...maintenance personnel or our customer service person nel Every 6 12 months or after each repair a comprehensive examination must be performed by trained and qualified technical service personnel inclu...

Страница 32: ...ative Before testing or maintenance clean and disinfect the device Inspection Maintenance Item Frequency Check the safety according to IEC 60601 1 At least once every two years after replacing the pow...

Страница 33: ...on equipment no explosion protection Liquid inlet protection grade IPX2 Operating mode Continuous mode Movement Hand held equipment 12 1 2 Power Power Adapter Input AC 100 240V 50 60 Hz Output DC 5V 2...

Страница 34: ...g them medium skin are 4 subjects light skin are 5 subjects dark skin are 3 sub jects the age from 21 to 28 Over the range of 70 to 100 overall accuracy was determined by calculating the root mean squ...

Страница 35: ...n the technical specifications it may result in inaccurate measure ments Mobile communication devices or wireless network devices may have an impact on the device 13 1 Electromagnetic emissions The BT...

Страница 36: ...d 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection fr...

Страница 37: ...equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 d 3 5 d 1 2 Resp d 3 5 80 to 800MHz d 1 2 800M to 2 5GHz where P is the maximum output power rating of the...

Страница 38: ...through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on Consumers noticed by Fair Trade Commission of Re...

Отзывы: