BT-36 Operation manual 8
P/N:36-ENG-OPM-EXP-R00
Bistos
2022.03
④
MEMORY
button
When the button is pressed, following icon
appears on
the screen and enters MEMORY mode.
The previously measured temperature values are saved up to
10 times.
⑤
POWER(START)
button
The button is for turn on the device or start temperature
measurement.
If the button is pressed while the device is turned off, the
power will be turned on. Whenever the button is pressed while
the device power is on, the temperature is measured through
the sensor.
(The device turns off if there is no any action for a certain
period of time.)
⑥
Temperature
sensing part
IR sensor that can measure body temperature or object
temperature without any contacts is placed in this part.
⑦
Battery cover
This device uses 2 AAA type batteries.
Press the top of the cover on the back of the device to open
it, insert the batteries and close the cover to use.
2.2.
LCD descriptions
Figure
Name
Description
Start screen
After the POWER is on, measure the ambient
temperature around the sensor once.
Ready screen Basic screen display after the start screen.
Body
temperature
measurement
screen
If you press the start button while this symbol
(body temperature measurement mode) is
displayed on the upper left of the LCD screen, the
blue light illuminates to confirm the measured body
temperature. If the measured value is higher than
37.5
℃
(99.5
℉
), the red light will illuminated to
indicate that user have fever.
Object
temperature
If you press the start button while this symbol