BT-36 Operation manual 15
P/N:36-ENG-OPM-EXP-R00
Bistos
2022.03
CISPR 11
suitable for use at home or in hospitals.
Harmonic
emissions
IEC 61000-3-2
Not
applicable
This device is powered by batteries only.
Voltage
fluctuations /
flicker
emissions
IEC 61000-3-3
Not
applicable
7.2. Electromagnetic immunity
BT-36 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BT-36 should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV Contact
±2 kV, ±4 kV, ±8 kV,
±15 kV Air
±8 kV HCP,
±8 kV VCP
Complies
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Immunity RF
IEC 61000-4-3
10V/m 80MHz -2.7GHz Complies
The electric field strength outside
the electromagnetic wave
shielding location from the fixed
RF transmitter determined
through the investigation of the
electromagnetic wave generation
location should be less than
3V/m.
Interference may occur in the
vicinity of equipment marked
with the following symbol.
Recommended
separation
distance
d = 1.2
P
150 kHz ~ 80 MHz
d = 1.2
P
80 MHz ~ 800 MHz
d = 2.3
P
800 MHz ~ 2.7 MHz
Where
P
is the maximum output
Conducted RF
IEC61000-4-6
3Vrms
150 kHz to 80 MHz
Not
applicable (No
electrical
cable
connection)