BT-36 Operation manual 16
P/N:36-ENG-OPM-EXP-R00
Bistos
2022.03
power rating of the
transmitter (W) according to the
transmitter manufacturer and
d
is the recommended separation
distance in meters(m).
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
Not
applicable
This device is powered by
batteries only.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not
applicable
Power frequency (50
Hz and 60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
Complies
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
> 95% dip for 0.5
cycles
60% dip for 5 cycles
70% dip for 25 cycles
95% dip for 5 seconds
Not
applicable
This device is powered by
batteries only.