Sequence BioTech,
LLC
28
•
Press the
CANCEL
button to exit the programming menu.
“Save Changes”
is displayed.
•
rotate the
Control wheel
to select
“Yes”
or
“No
.
”
Select
“Yes”
and
ENTER
to save the program into non-volatile memory.
Save Changes? Yes
Select and press Enter
To test the program:
•
Press
MENU
and then select
“Program
.
”
Menu: Program
Select and press enter
•
Press
ENTER
and then select
“Run
.
”
Program: Run
Select and press Enter
•
Press
ENTER
to run the program.
“Enter to continue”
appears
on the display screen until the user presses
ENTER
again. once
ENTER
is pressed, the program ends and the display returns
back to the main display screen.
Enter to continue
•
Press
ENTER
to return back to the main display screen.
Ready
Vol: 0 uL
standard Pipetting Program example
the following program contains multiple steps and is intended
to be used as a teaching tool to understand the step-based
programming method used with the Precision Series Pipettor.
A program containing up to 100 individual lines can be saved into
the internal flash memory. Keep in mind that when modifying an
existing program, it is often beneficial to use the
“Edit”
upper level
command (see page 14).
0 Repeat 5
10 Wait
1 Cls
11 Print Line1 “Dispense”
2 Print Line1 “Prepare to Aspirate” 12 Disp 15
3 Print Line2 “Enter to continue”
13 Next
4 Wait
14 Cls
5 Print Line1 “Aspirate”
15 Print Line1 “Done”
6 Asp 15
16 Print Line2 “Enter to continue”
7 Cls
17 Wait
8 Print Line1 “Prepare to Dispense” 18 End
9 Print Line2 “Enter to continue”