Table 3. Exchangeable block capacity in CH 3-150
Block
model
Description
Form of block
socket bottom
Catalogue number
B23-1.5
23 sockets for 1.5 ml microtubes
Conical
BS-010418-DK
B18-12
18 sockets Ø12 mm, depth 58 mm
Round
BS-010418-EK
B10-13
10 sockets Ø13 mm, depth 30 mm
Flat
BS-010418-LK
B10-16
10 sockets Ø16 mm, depth 56 mm
Flat
BS-010418-BK
B6-25
6 sockets Ø25 mm, depth 40 mm
Flat
BS-010418-CK
B5-29
5 sockets Ø29 mm, depth 40 mm
Flat
BS-010418-KK
B2-50
2
sockets Ø48 mm, depth 58 mm
Flat
BS-010418-AK
Biosan is committed to a continuous program of improvement and reserves the right
to alter design and specifications of the equipment without additional notice.
Maintenance
8.1.
If the unit requires maintenance electric circuit, disconnect the unit from electric circuit
and contact Biosan or your local Biosan representative.
8.2.
All maintenance and repair operations must be performed only by qualified and
specially trained personnel.
8.3.
Standard ethanol solution (75%) or other cleaning agents recommended for cleaning
of laboratory equipment can be used for cleaning and decontamination of the unit.
8.4.
Fuse replacement (model
CH 3-150
)
-
Disconnect from electric circuit.
-
Remove the power plug from the rear side of the unit.
-
Pull out the fuse holder by applying leverage in recess
(fig. 6/A).
-
Remove the fuse from the holder.
-
Check and replace with the correct fuse if necessary,
M
2
A for 230 V or
M
3.15 A for 120 V, (type
M
- time lag:
M
edium).
Fig. 6. Fuse replacement
8.5.
Error codes in case of a defect.
Some malfunctions trigger an error code to appear on display, accompanied by a
sound signal every 8 s. Press
RPM RUN/STOP
key to turn off the signal. Error code
format is letters
ER
and a number from
1
to
5
.
Disconnect the unit from the electric circuit and report the error code to Biosan or
your local Biosan representative.