8
Caution!
After the set time elapses and the timer stops, the device does not
stop the heating or cooling process. Press the
RUN/STOP T (ºC)
key
(fig. 1/7) manually to stop the heating or cooling.
5.1.11. The timer can be stopped before the set time elapses if required by pressing the
RUN/STOP TIME
key. Pressing the key again restarts the timer.
5.1.12. The set time value can be changed during operation without stopping the timer.
5.1.13. After finishing the operation, turn off (position
O
) the unit switching off the power
switch on the rear panel. Disconnect the external power supply from electric circuit.
5.2.
Working with model
CH 3-150
.
Fig. 2. Control panel of CH 3-150
5.2.1. Connect the power cord to a grounded power socket and switch on (position
I
) the
power switch located on the rear panel of the unit.
5.2.2. The unit will switch on and the following readouts will be shown on the display:
-
In the upper line
Set p.
: previously set time and temperature parameters;
-
In the lower line
Actual p.
: program indication STOP and current temperature.
Caution!
Temperature starts changing automatically after switching on
according to selected settings. Time of temperature stabilization
depends on the initial temperature.
5.2.3. The unit has 16 cooling and 16 heating programs for temperature and time setup.
One cooling and three heating programs are factory pre-set. The programs have
following settings:
Program
Temperature Time, hh:mm
Program
Temperature Time, hh:mm
Cooling [P1]
-
3 °C
1:00
Heating [P1]
+80 °C
0:30
Cooling [P2-16]
-
3 °C
0:00
Heating [P2]
+105 °C
0:07
Heating [P3]
+150 °C
2:00
Heating [P4-16]
+25 °C
0:00
Содержание CH 3-150
Страница 1: ......