12
Specification
The unit is designed for operation in cold rooms, incubators (except CO
2
incubators)
and laboratory rooms at ambient temperature from +4C to +40°C in a non-condensing
atmosphere and maximum relative humidity 8
0% for temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C.
Temperature specifications
CH-100
CH 3-150
Setting range
Cooling
-
10°C … +100°C
-
3°C … +20°C
Heating
+25°C … +150°C
Control range
Cooling
30°C below RT*… +100°C
23°C below RT …
5°C below RT
Heating
5°C above RT … +150°C
Setting resolution
Cooling
±0.1°C
±0.1 °C
Heating
±1 °C
Stability
±0.1°C
Uniformity at +37°C
±0.1°C
Heat up time from 25°C to 100°C
16 min
12 min
Cool down time from 25°C
21 min (to -
10°C)
60 min (to -
3°C)
Temperature calibration option
yes
Calibration
coefficient range
Heating
0.936 … 1.063 (±0.063)
0.936 … 1.063 (±0.063)
Cooling
-
General specifications
CH-100
CH 3-150
Digital time setting
1 min
– 96 h
1 min
– 99 h 59 min
Digital time setting increment
1 min
LCD Display
2x16 symbols
2x24 symbols
Dimensions
240x260x165 mm
295x285x220 mm
Weight**
3.2 kg
5.6 kg
Power consumption
60 W
430 W
Input current
DC 12 V, 5.0 A
AC 230 V, 50 Hz, 1.9 A or
AC 120 V, 60 Hz, 3.6 A
External power supply
In AC 100-240 V, 50/60 Hz
Out DC 12 V
-
Table 2. Built-in block capacity in CH-100
Built-in block
Capacity
Catalogue number
CH-1
20 x 0.5 ml and 12 x 1.5 ml microtubes
BS-010410-BAI
CH-2
20 x 1.5 ml microtubes
BS-010410-BAI
CH-3
20 x 2 ml microtubes
BS-010410-BAI
*
Room temperature
**
Accurate within
±10%.