  Specification 

The unit is designed for operation in cold rooms, incubators (except CO

2

 incubators) 

and  laboratory  rooms  at  ambient  temperature  from  +4C  to  +40°C  in  a  non-condensing 
atmosphere and maximum relative humidity 8

0% for temperatures up to 31°C decreasing 

linearly to 50% relative humidity at 40°C. 

Temperature specifications 

CH-100 

CH 3-150 

Setting range 

Cooling 

-

10°C … +100°C 

-

3°C … +20°C 

Heating 

+25°C … +150°C 

Control range 

Cooling 

30°C below RT*… +100°C 

23°C below RT …  

5°C below RT 

Heating 

5°C above RT … +150°C 

Setting resolution 

Cooling 

±0.1°C 

±0.1 °C 

Heating 

±1 °C 

Stability 

±0.1°C 

Uniformity at +37°C 

±0.1°C 

Heat up time from 25°C to 100°C 

16 min 

12 min 

Cool down time from 25°C 

21 min (to -

10°C) 

60 min (to -

3°C) 

Temperature calibration option 

yes 

Calibration 

coefficient range 

Heating 

0.936 … 1.063 (±0.063) 

0.936 … 1.063 (±0.063) 

Cooling 

 

General specifications 

CH-100 

CH 3-150 

Digital time setting 

1 min 

– 96 h 

1 min 

– 99 h 59 min 

Digital time setting increment 

1 min 

LCD Display 

2x16 symbols 

2x24 symbols 

Dimensions 

240x260x165 mm 

295x285x220 mm 

Weight** 

3.2 kg 

5.6 kg 

Power consumption 

60 W 

430 W 

Input current  

DC 12 V, 5.0 A  

AC 230 V, 50 Hz, 1.9 A or 

AC 120 V, 60 Hz, 3.6 A 

External power supply 

In AC 100-240 V, 50/60 Hz 

Out DC 12 V 

 

Table 2. Built-in block capacity in CH-100 

Built-in block 

Capacity 

Catalogue number 

CH-1 

20 x 0.5 ml and 12 x 1.5 ml microtubes 

BS-010410-BAI 

CH-2 

20 x 1.5 ml microtubes 

BS-010410-BAI 

CH-3 

20 x 2 ml microtubes 

BS-010410-BAI 

 

 

 

 

Room temperature 

** 

Accurate within 

±10%. 

Содержание CH 3-150

Страница 1: ......

Страница 2: ...peration 7 Calibration 10 Specification 12 Maintenance 13 Warranty and Claims Registration 14 EU Declaration of conformity 15 About this edition of the operating manual The manual applies to following model and versions of heating cooling system CH 100 versions V 7A01 V 7A02 and V 7A03 CH 3 150 versions V 2AD and V 2AE ...

Страница 3: ...e only the external power supply CH 100 provided with the unit Do not use an ungrounded power socket or an ungrounded extension lead Ensure that the switch and the plug are easily accessible during use If liquid penetrates into the unit disconnect it from electric circuit and have it checked by a repair and maintenance technician Disconnect the unit from electric circuit before moving Do not opera...

Страница 4: ...ntrol panel is divided in two parts for separate control of cooling and heating blocks Blocks are controlled independently with possibility of setting up to 16 custom programs for different temperature and thermostating time Seven thermoblocks are available for different size and quantity of tubes Models can be used in Molecular and cell biology for nucleic acid extraction and analysis Protein ana...

Страница 5: ...in the original package 4 2 Complete set Package contains 4 2 1 CH 3 150 Combitherm 2 CH 3 150 Combitherm 2 heating cooling dry block system 1 pce Transparent block lids 2 pcs Spare fuse inside fuse holder 1 pce Power cable 1 pce Operating manual declaration of conformity 1 copy B23 1 5 block on request B18 12 block on request B10 13 block on request B10 16 block on request B6 25 block on request ...

Страница 6: ...onnect the external power supply and the power cable Plug the external power supply into the socket on the rear side of the unit and position the unit with easy access to the power switch and plug 4 4 Heating or cooling block installation and replacement model CH 3 150 Caution Heating cooling block surface may become very hot or cold during the operation Avoid touching the block or heating cooling...

Страница 7: ...lay upper line fig 1 3 Temperature incre ment is 0 1ºC Pressing the key down for more than 2 s makes the values change quicker 5 1 4 Press the RUN STOP T ºC key fig 1 8 to start heating or cooling The corres ponding operation status is indicated on the display H for heating C for cooling fig 1 1 Current temperature value is shown on display lower line fig 1 5 5 1 5 To stop the heating or cooling p...

Страница 8: ...ord to a grounded power socket and switch on position I the power switch located on the rear panel of the unit 5 2 2 The unit will switch on and the following readouts will be shown on the display In the upper line Set p previously set time and temperature parameters In the lower line Actual p program indication STOP and current temperature Caution Temperature starts changing automatically after s...

Страница 9: ...3 Use the and Time keys fig 2 10 to set the required work time interval in hours and minutes Time increment is 1 minute Set values are shown on display on the right in the upper line Set p fig 2 12 Press the PH key to save settings and return to work mode 5 2 6 Program execution When the necessary program is selected with PC or PH key heating or cooling processes start automatically 5 2 7 After th...

Страница 10: ...rature readings equal leave the unit for 30 min for thermal stabilization 6 4 4 Let us assume that the readings of independent sensor is 39 C but the display s actual temperature is 40 C fig 3 4 Then it is necessary to add 1 C correction 6 4 5 Hold the RUN STOP TIME or Start Stop key pressed for more than 8 s to activate calibration mode The following parameters will be shown on the display fig 3 ...

Страница 11: ...11 Fig 4 Control panel in calibration and operation mode Fig 5 Control panel in operation mode after calibration ...

Страница 12: ...6 min 12 min Cool down time from 25 C 21 min to 10 C 60 min to 3 C Temperature calibration option yes Calibration coefficient range Heating 0 936 1 063 0 063 0 936 1 063 0 063 Cooling General specifications CH 100 CH 3 150 Digital time setting 1 min 96 h 1 min 99 h 59 min Digital time setting increment 1 min LCD Display 2x16 symbols 2x24 symbols Dimensions 240x260x165 mm 295x285x220 mm Weight 3 2 ...

Страница 13: ... Biosan representative 8 2 All maintenance and repair operations must be performed only by qualified and specially trained personnel 8 3 Standard ethanol solution 75 or other cleaning agents recommended for cleaning of laboratory equipment can be used for cleaning and decontamination of the unit 8 4 Fuse replacement model CH 3 150 Disconnect from electric circuit Remove the power plug from the rea...

Страница 14: ... warranty is available free of charge after registration during 6 months from the date of sale Online registration form can be found in section Warranty registration on our website at the link below For CH 100 a Basic Plus class model extended warranty is a paid service Contact your local Biosan representative or our service department through the Technical support section on our website at the li...

Страница 15: ...tandards LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General requirements LVS EN 61010 1 2011 Safety requirements for electrical equipment for measurement control and laboratory use General requirements LVS EN 61010 2 010 2015 Particular requirements for laboratory equipment for the heating of materials We declare that this product conforms ...

Страница 16: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 2 7 01 February 2017 ...

Отзывы: