background image

15

If you continue with the same tip

a.  Change the volume by pressing 

  

b. Con

fi

 rm your selection by pressing 

 

c.  Change the number of aliquots by pressing 

  

d. Con

fi

 rm your selection by pressing 

  

If you want to change the tip size or tip, go to  step 1.
If you want to eject the tip, press 

Note:

  It  is  possible  to  interrupt  the  dispensing  by  pressing  the  ENTER 

button 

. The display shows E and the tip can be emptied by pressing the  

START button 

.

Note: 

When aspirating the liquid using the 25 or 50 ml tip, let the tip stay in 

the liquid a few seconds after the plunger has stopped moving. This way you 

will ensure that all liquid has been aspirated into the tip.

Note:

 It is possible to work with aqueous liquids and liquids with higher 

viscosity. However, it is recommended always to choose the tip with the 

smallest applicable volume in conjunction with high viscosity liquids. Please 

note, the speci

fi

 cations are given in accordance with ISO8655-5.

6. MAINTENANCE

6.1. Cleaning the dispenser

To  ensure  trouble-free  operation  it  is  recommended  to  apply  regular 

maintenance and cleaning to Biohit 

e

LINE Lite Dispenser.

To clean and decontaminate the outer surface use Proline Biocontrol (Cat. 

No. 724004) and a soft lint-free cloth. Gently clean the surface of the pipette 

with the moistened cloth and wipe dry. It is possible to use ethanol (70%) 

isopropanol (60%) or mild detergent as a cleaning agent.

Note:

 Remove the battery pack from the dispenser prior the cleaning! 

It is recommended always to use gloves when cleaning the dispenser. Do not 

let the liquid enter the pipette handle!

Содержание eLINE Lite

Страница 1: ...Instruction Manual eLINE Dispenser Dispenser Lite Lite...

Страница 2: ...HARGING OF THE DISPENSER 8 4 1 Charging with the Biohit eLINE Charging Stand and Charging Carousel 9 4 2 Charging through direct charging system 10 4 3 Charging Recommentations 10 4 4 Electrical speci...

Страница 3: ...ior construction of the Biohit eLINE Lite Dispenser enable maximum liquid handling performance with high levels of accuracy and precision The eLINE Lite Dispenser uses positive displacement principle...

Страница 4: ...Starter Kit includes 0 5 ml 20 pcs 1 0 ml 20 pcs 2 5 ml 20 pcs 5 0 ml 20 pcs 10 0 ml 20 pcs 1 adapter included Note The tips are disposable do not reuse them 3 QUICK START 3 1 Unpacking The Biohit eL...

Страница 5: ...lastic protector Fig 1 1 Push the charging buttons simultaneuously in order to open the battery cover 2 Remove the plastic protector Note It is recommended to charge the pipette for 12 hours before in...

Страница 6: ...of the Biohit eLINE Lite Dispenser is performed using the display keyboard operating buttons and tip Fig 2 and 3 Keyboard Display Start button Tip ejection buttons Tip connector Biohit Dispenser Tip...

Страница 7: ...ple dispensing mode with Tip Guide The Tip Guide is a unique built in guide that selects the optimal tip size according to selected volume and number of dispensings 3 3 3 Keyboard and operation button...

Страница 8: ...cted operating mode Only a soft touch is required to operate the button 3 3 6 Tip Attachment and Ejection Tip attachment 1 Push the tip into the tip connector Make sure that the tip is properly attach...

Страница 9: ...e dispenser Please note that if enter is not pressed the dispenser will not start charging even placed in charging stand connected to AC adaptor Note In case you continue to work with the dispenser af...

Страница 10: ...adaptor to the electrical outlet check that its voltage setting corresponds to the voltage setting of your electrical outlet Note The Charging Carousel is used to charge the dispenser without the tip...

Страница 11: ...battery cover is properly closed 3 Connect the AC adaptor to the electrical outlet as well as to the socket at the back of the dispenser 4 The charging time is 6 hours for the completely discharged b...

Страница 12: ...ltages and main plug according to local requirements Output voltage 9 VDC 360 mA AC adaptor for Charging Carousel Input voltages and main plug according to local requirements Output voltage 9 VDC 1200...

Страница 13: ...uots Select the tip size Tip Guide Shows the best tip option If there are also other tip options available the TIP SIZE text is blinking View the other options by pressing the Arrow up down key Do you...

Страница 14: ...t is attached 2 Select the desired aliquot volume by pressing Note When holding down the the volume display start to scroll faster 3 Press to confirm the selection 4 Select the number of aliquots by p...

Страница 15: ...tip above the receiving vessel and press START button and repeat until the cycle is completed 17 Finally position the tip to discard excess volume and click twice the START button 18 Now the Dispenser...

Страница 16: ...ith aqueous liquids and liquids with higher viscosity However it is recommended always to choose the tip with the smallest applicable volume in conjunction with high viscosity liquids Please note the...

Страница 17: ...ufacturer The use of any other type of battery will immediately void the warranty 7 TROUBLESHOOTING The Biohit eLINE Lite Dispenser have a onboard monitoring program to control the performance of each...

Страница 18: ...se original tips Clean the tip connector and attach new tip Dispenser out of the established specifications Unsuitable tip Instrument damaged Use original tips Return to your Biohit service representa...

Страница 19: ...ENANCE A OF SERVICE OR NEGLIGENCE OF REGULAR MAINTENANCE AND SERVICE ACCIDENTAL DAMAGE INCORRECT STORAGE OS USE OF THE PRODUCTS FOR OPERATIONS OUTSIDE THEIR SPECIFICATIONS CONTRARY TO THE INSTRUCTIONS...

Страница 20: ...10 1 00 1 00 0 50 2 00 1 100 0 2 2 200 0 4 200 20 1 00 1 70 0 50 2 50 1 100 0 5 5 500 1 500 50 0 80 1 00 0 60 3 00 1 100 1 0 10 1000 2 1000 100 0 50 0 50 0 20 0 60 1 100 2 5 25 2500 5 2500 250 0 60 0...

Страница 21: ...ectrical and electronic equipments this device must not be recycled as unsorted municipal waste Instead this device must be collected separately in accordance the local recycling regulations The solid...

Страница 22: ...www biohit com Due to Biohit s continuing R D effort specifications may change without prior notice Biohit Proline innovations are covered by multiple patents including U S Pat 5 343 769 U S Pat 5 50...

Отзывы: