15
If you continue with the same tip
a. Change the volume by pressing
b. Con
fi
rm your selection by pressing
c. Change the number of aliquots by pressing
d. Con
fi
rm your selection by pressing
If you want to change the tip size or tip, go to step 1.
If you want to eject the tip, press
Note:
It is possible to interrupt the dispensing by pressing the ENTER
button
. The display shows E and the tip can be emptied by pressing the
START button
.
Note:
When aspirating the liquid using the 25 or 50 ml tip, let the tip stay in
the liquid a few seconds after the plunger has stopped moving. This way you
will ensure that all liquid has been aspirated into the tip.
Note:
It is possible to work with aqueous liquids and liquids with higher
viscosity. However, it is recommended always to choose the tip with the
smallest applicable volume in conjunction with high viscosity liquids. Please
note, the speci
fi
cations are given in accordance with ISO8655-5.
6. MAINTENANCE
6.1. Cleaning the dispenser
To ensure trouble-free operation it is recommended to apply regular
maintenance and cleaning to Biohit
e
LINE Lite Dispenser.
To clean and decontaminate the outer surface use Proline Biocontrol (Cat.
No. 724004) and a soft lint-free cloth. Gently clean the surface of the pipette
with the moistened cloth and wipe dry. It is possible to use ethanol (70%)
isopropanol (60%) or mild detergent as a cleaning agent.
Note:
Remove the battery pack from the dispenser prior the cleaning!
It is recommended always to use gloves when cleaning the dispenser. Do not
let the liquid enter the pipette handle!