3.3.4. Reset 

Button

To reset the dispenser, press the reset button on the battery cover. The 

display shows 

E

 and 

TIP

 in blinking form and pressing the tip ejector button 

will reset the pipette to the 

d

-mode.

Note: 

The dispenser should never be reset 

during pipetting! Resetting will cause the 

plunger to move to the home position and it 

will empty the contents of the tip. The tip is 

also ejected in the end.

3.3.5. Start 

Button

The start button triggers the aspiration and 

dispensing  operations  according  to  the 

selected operating  mode. Only a soft touch 

is required to operate the button. 

3.3.6.   Tip Attachment and  

             Ejection

                                                                                         

Tip attachment

1.  Push the tip into the tip connector. 

Make 

sure that the tip is properly attached. 

(Fig. 4.)

2. Con

fi

 rm the tip size by double clicking the 

Start-button according the instruction on 

the page 15.

Releasing the tip:

1.  Press the Tip Ejection button. The tip is 

released immediately (Fig. 5.)

Note:

 After the dispensing cycle has been 

completed a small residual amount of 

fl

 uid may have been left in the tip. During 

the tip ejection it is possible that a part 

of the remaining 

fl

 uid comes out of the 

tip.  Therefore  it  is  recommended  that  the 

tip ejection is performed above a waste 

container.

1

1

2

Fig. 4.

Fig. 5.

Содержание eLINE Lite

Страница 1: ...Instruction Manual eLINE Dispenser Dispenser Lite Lite...

Страница 2: ...HARGING OF THE DISPENSER 8 4 1 Charging with the Biohit eLINE Charging Stand and Charging Carousel 9 4 2 Charging through direct charging system 10 4 3 Charging Recommentations 10 4 4 Electrical speci...

Страница 3: ...ior construction of the Biohit eLINE Lite Dispenser enable maximum liquid handling performance with high levels of accuracy and precision The eLINE Lite Dispenser uses positive displacement principle...

Страница 4: ...Starter Kit includes 0 5 ml 20 pcs 1 0 ml 20 pcs 2 5 ml 20 pcs 5 0 ml 20 pcs 10 0 ml 20 pcs 1 adapter included Note The tips are disposable do not reuse them 3 QUICK START 3 1 Unpacking The Biohit eL...

Страница 5: ...lastic protector Fig 1 1 Push the charging buttons simultaneuously in order to open the battery cover 2 Remove the plastic protector Note It is recommended to charge the pipette for 12 hours before in...

Страница 6: ...of the Biohit eLINE Lite Dispenser is performed using the display keyboard operating buttons and tip Fig 2 and 3 Keyboard Display Start button Tip ejection buttons Tip connector Biohit Dispenser Tip...

Страница 7: ...ple dispensing mode with Tip Guide The Tip Guide is a unique built in guide that selects the optimal tip size according to selected volume and number of dispensings 3 3 3 Keyboard and operation button...

Страница 8: ...cted operating mode Only a soft touch is required to operate the button 3 3 6 Tip Attachment and Ejection Tip attachment 1 Push the tip into the tip connector Make sure that the tip is properly attach...

Страница 9: ...e dispenser Please note that if enter is not pressed the dispenser will not start charging even placed in charging stand connected to AC adaptor Note In case you continue to work with the dispenser af...

Страница 10: ...adaptor to the electrical outlet check that its voltage setting corresponds to the voltage setting of your electrical outlet Note The Charging Carousel is used to charge the dispenser without the tip...

Страница 11: ...battery cover is properly closed 3 Connect the AC adaptor to the electrical outlet as well as to the socket at the back of the dispenser 4 The charging time is 6 hours for the completely discharged b...

Страница 12: ...ltages and main plug according to local requirements Output voltage 9 VDC 360 mA AC adaptor for Charging Carousel Input voltages and main plug according to local requirements Output voltage 9 VDC 1200...

Страница 13: ...uots Select the tip size Tip Guide Shows the best tip option If there are also other tip options available the TIP SIZE text is blinking View the other options by pressing the Arrow up down key Do you...

Страница 14: ...t is attached 2 Select the desired aliquot volume by pressing Note When holding down the the volume display start to scroll faster 3 Press to confirm the selection 4 Select the number of aliquots by p...

Страница 15: ...tip above the receiving vessel and press START button and repeat until the cycle is completed 17 Finally position the tip to discard excess volume and click twice the START button 18 Now the Dispenser...

Страница 16: ...ith aqueous liquids and liquids with higher viscosity However it is recommended always to choose the tip with the smallest applicable volume in conjunction with high viscosity liquids Please note the...

Страница 17: ...ufacturer The use of any other type of battery will immediately void the warranty 7 TROUBLESHOOTING The Biohit eLINE Lite Dispenser have a onboard monitoring program to control the performance of each...

Страница 18: ...se original tips Clean the tip connector and attach new tip Dispenser out of the established specifications Unsuitable tip Instrument damaged Use original tips Return to your Biohit service representa...

Страница 19: ...ENANCE A OF SERVICE OR NEGLIGENCE OF REGULAR MAINTENANCE AND SERVICE ACCIDENTAL DAMAGE INCORRECT STORAGE OS USE OF THE PRODUCTS FOR OPERATIONS OUTSIDE THEIR SPECIFICATIONS CONTRARY TO THE INSTRUCTIONS...

Страница 20: ...10 1 00 1 00 0 50 2 00 1 100 0 2 2 200 0 4 200 20 1 00 1 70 0 50 2 50 1 100 0 5 5 500 1 500 50 0 80 1 00 0 60 3 00 1 100 1 0 10 1000 2 1000 100 0 50 0 50 0 20 0 60 1 100 2 5 25 2500 5 2500 250 0 60 0...

Страница 21: ...ectrical and electronic equipments this device must not be recycled as unsorted municipal waste Instead this device must be collected separately in accordance the local recycling regulations The solid...

Страница 22: ...www biohit com Due to Biohit s continuing R D effort specifications may change without prior notice Biohit Proline innovations are covered by multiple patents including U S Pat 5 343 769 U S Pat 5 50...

Отзывы: