19
9. PERFORMANCE
SPECIFICATIONS
Note:
The manufacturer’s speci
fi
cations should be used as guidelines when
establishing your own performance speci
fi
cations accordance with ISO
8655-5.
eLINE Lite Dispenser and Biohit Dispenser Tip system
Tip
volume
(ml)
Volume
min/
max (
μ
l)
Increment
(
μ
l)
Test
volume
(
μ
l)
Inacc.
(%)
Impr.
(%)
Number
of
dispensing
min/max
0.1
1/100
0.2 100
10
1.00
1.00
0.50
2.00
1/100
0.2
2/200
0.4
200
20
1.00
1.70
0.50
2.50
1/100
0.5
5/500
1
500
50
0.80
1.00
0.60
3.00
1/100
1.0
10/1000
2
1000
100
0.50
0.50
0.20
0.60
1/100
2.5
25/2500
5
2500
250
0.60
0.60
0.20
0.40
1/100
5.0
50/5000
10
5000
500
0.60
0.60
0.15
0.50
1/100
10.0
100/10 000
20
10 000
1000
0.40
0.40
0.20
1.00
1/100
25.0
500/25 000
50
25 000
2500
0.50
0.50
0.15
0.50
1/50
50.0
1000/50 000
100
50 000
5000
0.30
0.30
0.15
0.70
1/50
Note:
The speci
fi
cations are valid only to the
e
LINE Lite Dispenser and Biohit
e