PowerPac Universal Power Supply Instruction Manual
Page | 27
Error Message
Cause
Solution
ERROR STOP:
CODE 14
V LIMIT
RUN PAUSED
CONTINUE CANCEL
Voltage limit exceeded Press the CONTINUE soft key to
continue the run. If the problem
persists, contact Bio-Rad Technical
Support.
ERROR STOP:
CODE 15
A LIMIT
RUN PAUSED
CONTINUE CANCEL
Amperage limit
exceeded
Press the CONTINUE soft key to
continue the run. If the problem
persists, contact Bio-Rad Technical
Support.
ERROR STOP:
CODE 16
W LIMIT
RUN PAUSED
CONTINUE CANCEL
Wattage limit exceeded Press the CONTINUE soft key to
continue the run. If the problem
persists, contact Bio-Rad Technical
Support.
ERROR STOP:
CODE 17
OVERHEATING
Internal maximum
temperature exceeded –
potential causes:
blocked vent, fan failure,
component failure
Check for obstructions to fan opening
and/or ventilation slots. Check that
fan is operating. If unable to resolve,
contact Bio-Rad Technical Support.
ERROR STOP:
CODE 18
OPERATING BELOW 10V
Operating below 10 V
(connected load and
parameter values would
require operating < 10 V
output)
Not applicable to voltage mode.
If operating in current mode, increase
the current (mA) setting or decrease
the load (increase its resistance).
If operating in wattage mode,
increase the power (W) setting or
decrease the load resistance.
Содержание PowerPac Universal Power Supply
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 12: ...Page 4 PowerPac Universal Power Supply Instruction Manual ...
Страница 14: ...Page 6 PowerPac Universal Power Supply Instruction Manual ...
Страница 30: ...Page 22 PowerPac Universal Power Supply Instruction Manual ...
Страница 38: ...Page 30 PowerPac Universal Power Supply Instruction Manual ...
Страница 40: ...Page 32 PowerPac Universal Power Supply Instruction Manual ...
Страница 41: ......