![BIO RAD Gel Doc 2000ChemiDocChemiDoc XRS Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/bio-rad/gel-doc-2000chemidocchemidoc-xrs/gel-doc-2000chemidocchemidoc-xrs_hardware-instruction-manual_2753168037.webp)
Appendix B: ChemiDoc
System installation
The following appendix gives additional information to use the Universal Hood in the ChemiDoc
configuration.
Connecting the cabling harnesses
a) Carefully take the CCD camera assembly out of the box and hold the camera so that the two
locking holes are facing you.
b) Insert it into the camera holder from the open left side and attach it with the two black screws.
c) Place the bracket with the camera on the top of the Hood and attach it with the four washers and
knobs.
d) Connect each end of the CCD Camera cable following the instruction below:
Label / Color
Instrument / Connection
9-pin connector / Black
PCI board / Back of computer
BNC Black
Hood rear left side / Video
Multi Pin Red
Hood rear left side / Int
Multi pin / Metal end
CCD Camera / 12 V AC-DC
Multi pin / Metal end
Hood rear left side/ Camera
Not used with the
ChemiDoc CCD Camera
Serial port DB25 to connect
a null modem serial cable
from PC Com1 port. This
option permits to control the
MZL directly from a
program
on the PC.
Multi pin from MZL
Multi pin from PCI board
BNC end from PCI board
Multi pin from CCD Camera
Page 37 of 42