Picture of the Sliding Filter from the inside
4.3.2 Initial test
Before starting the initial test please make sure that:
1. The software and the digitizing board are installed properly.
2. The cabling harness is connected properly. (See
Appendix A,
B or C
)
Please follow the procedure in the table to ensure that the Universal Hood is functioning
properly.
Initial test for the Universal Hood
Procedure
Control Panel
Make sure the door and drawer are closed and the PC is switched on.
Turn on the system.
Power LED turns on after short blinking.
Press the Epi-White key.
Epi-White LED turns on.
Open the door.
Epi-lights are on, Tran UV LED blinks.
Press the Epi-White key again.
Close the door.
Epi-lights and LED are turned off.
Trans UV LED turns off.
Press the Trans UV key.
Trans UV LED turns on.
Open the door.
Trans UV LED blinks.
Close the door.
Trans UV LED turns off.
Press the Trans UV key.
Trans UV LED turns on.
Open the drawer.
Trans UV LED blinks.
Close the drawer.
Trans UV LED turns off.
Press the Trans UV key.
Trans UV LED turns on.
Press the Prep UV key
Prep UV LED turns on.
Press the Hold key.
Hold LED turns on.
Open the drawer.
Trans UV LED blinks, Prep UV LED and
Hold LED turn off.
Close the drawer.
Trans UV LED turns off.
Note: When the Imaging LED blinks (PC switched Off or during the Integration) the
buttons of the Light Source section on the control panel are disabled.
Page 22 of 42