Connect the Camera Controller Cable to the PCI
card
4.2.4 Installing the software
Please refer to your software manual and release notes for proper software installation.
4.2.5 Connecting the cabling harnesses
The connections are different if you are installing a Gel Doc 2000, ChemiDoc or a ChemiDoc
XRS camera. Please refer to the instructions included in
Appendix A, B, C
(respectively) for
Gel Doc 2000/ChemiDoc/ChemiDoc XRS cable wiring.
The Controls for the Motorized Zoom Lens can be operated from the membrane touch pad of the
Universal Hood or from the PC using the buttons that appear in the window when the option Gel
Doc 2000, ChemiDoc or ChemiDoc XRS is selected in the start screen of the TDS Quantity One
program. To use this option it is necessary to connect the Universal Hood Serial port (DB25) on
the back left side to the COM1 of the PC. The PC/MAC cable is included.
NOTE: In case when a MAC has a USB port instead of the Serial Port, use a USB to Serial
Adapter (170-7959) to control the Lens via software
4.2.6 Installing the White Light Transilluminator (Optional)
The Universal Hood has two metal hooks on which the White Light Transilluminator can stand
up when not in use. When WL Tran illuminated images need to be acquired the White Light
Transilluminator can be positioned horizontally on the Trans UV surface. Refer to the below
picture to locate the hooks and the power cable that has to be connected to the White Light
Transilluminator. Remove the black rubber that covers the banana plug and insert it into the WL
outlet. Be sure that the WL main switch is in the on position. The WL Trans on/off is then
driven by the membrane touch pad.
Power
cable
Left
hook
Right
hook
Hooks and power cable of the WL Transilluminator