NCR-80 Four-wire 4-20 mA/HART
925-0332 Rev B
6 Set up the sensor with the display and adjustment module
35
Approx. 60 minutes after the last pressing of a key, an automatic reset to measured value indi-
cation is triggered. Any values not confirmed with
[OK]
will not be saved.
6.3 Measured value indication – Selection national lan-
guage
With the
[
-
>]
key you move between three different indication modes.
In the first view, the selected measured value is displayed in large digits.
In the second view, the selected measured value and a corresponding bar graph presentation
are displayed.
In the third view, the selected measured value as well as a second selectable value, e.g. the
temperature of the electronics, are displayed.
During the initial setup of an instrument shipped Ex works, use the "
OK
" key to get to the menu
"
National language
".
This menu item is used to select the national language for further parameter adjustment. You
can change the selection via the menu item "
Setup - Display, Menu language
".
With the "
OK
" key you move to the main menu.
6.4 Parameter adjustment - Quick setup
To quickly and easily adapt the sensor to the application, select the menu item "
Quick setup
" in
the start graphic on the display and adjustment module.
Select the individual steps with the
[
-
>]
key.
After the last step, "
Quick setup terminated successfully
" is displayed briefly.
Information:
The echo curve of setup is stored automatically during the quick setup.
The return to the measured value indication is carried out through the
[->]
or
[ESC]
keys or au-
tomatically after 3 s.
You can find "Extended adjustment" in the next sub-chapter.