14. Alarm functions
14.1 Alarm messages
WARNING: If customer should use a chamber running in non-supervised continuous opera-
tion, we strongly recommend in case of inclusion of irrecoverable specimen or samples to
split such specimen or samples and store them in at least two chambers, if this is feasible.
In the event of equipment failures, when the temperature deviates from the set tolerance range limits,
optical and possibly acoustic alarm messages are given out via the controller. A zero-voltage relay alarm
output (9) (chap. 14.4.7) permits transmission of the alarm e.g., to a central monitoring system.
The alarm messages door open and leaving the tolerance range are emitted after a configurable delay
(chap. 13), the others immediately when the fault occurs. The tolerance range alarm is suppressed after
opening the chamber door or turning on the freezer until the setpoint is reached and then for the selected
delay time.
In Normal display a text message indicates the alarm cause. The “collective alarm” icon flashes. If
the audible arm is activated, the buzzer sounds.
If more than one alarm signal is sent simultaneously, they are displayed in a cycle.
Press the
OK button
to confirm the alarm and mute the buzzer. If the alarm cause is still valid, the “collec-
tive alarm” icon is lit.
Alarm indication (example: safety controller alarm)
Safety controller
Alarm messages overview:
Condition
Alarm message
Moment of Alarm message and
switching the zero-voltage relay
alarm output
Chamber door open
“Door open”
after configurable time (chap. 13).
Factory setting: 1 minute
Exceeded setpoint of the safety
controller
“Safety controller”
immediately
The current actual temperature
value is outside the tolerance range
(chap. 13)
“Temp. range”
after configurable time (chap. 13).
Factory setting: 60 minutes
Continuous compressor operation,
refrigeration system defective. Con-
tact BINDER Service.
“Continuous operation”
•
12 hours after start-up, if the set-
point wasn’t reached
•
60 minutes after reaching the set-
point
Clean / replace the condenser air
filter (chap. 24.4.1)
“Condenser temp.”
immediately
Compressor defective. Contact
BINDER Service.
“Compressor defective”
immediately
Inner temperature sensor defective.
Control continues using the safety
controller temperature sensor
“Inner temp. sensor”
immediately
Содержание ULTRA.Guard UF V 500
Страница 106: ...UF V E3 09 2018 Page 106 116 27 7 Dimensions UF V 500 E3 Dimensions in mm...
Страница 107: ...UF V E3 09 2018 Page 107 116 27 8 Dimensions UF V 700 E3 Dimensions in mm...
Страница 108: ...UF V E3 09 2018 Page 108 116 28 Certificates and declarations of conformity 28 1 EU Declaration of Conformity...
Страница 109: ...UF V E3 09 2018 Page 109 116...
Страница 110: ...UF V E3 09 2018 Page 110 116 29 Product registration...
Страница 117: ......