UF V (E3) 09/2018
Page 52/116
13. Tolerance range and alarm delay settings
In this menu you can define the deviation between the actual value and setpoint of, which that shall cause
a tolerance range alarm. The entered value defines the limit of permitted deviations from the set-point
(exceeding and falling below). Reaching this limit triggers tolerance alarm.
In addition you can specify a delay time for the tolerance range alarm and the door open alarm.
During the delay time there is no alarm message after leaving the tolerance range:
•
After door opening: fixed delay time of 1 hour
•
After turning on the chamber or returning power supply: fixed delay time of 12 hours
•
After leaving the tolerance range in normal operation: adjustable delay time (chap. 13.2)
If the actual value is outside the tolerance range, after the configured alarm delay time the alarm message
“Temperature range” is displayed in Normal display (chap. 14.1). If the alarm buzzer is activated (chap.
14.3) there is an audible alert.
This function only activates after the set-point has been reached once.
Required access level: “Admin”.
13.1 Setting the delay time for door open alarm
Path:
Normal display
Settings
Various
Door alarm delay (min)
Press the
OK button
to enable the setting.
Setting the door alarm delay.
The current setting flashes. Use the
arrow buttons
to enter the desired
time after which the door open alarm shall be triggered. Entry range: 1
up to 300 minutes. Factory setting: 1 minute.
Confirm the entry with the
OK button
.
Door alarm delay (min
With the
arrow-down button
you can now change to the tolerance range alarm delay setting.
With the
Back button
you can go back to the “
Various
” submenu and, repeatedly pressing it, to
Normal
display
.
13.2 Setting the delay time for tolerance range alarm
Following opening the door or chamber startup, the tolerance range alarm becomes valid only after the
setpoint has been reached.
Path:
Normal display
Settings
Various
Range alarm delay
(min)
Press the
OK button
to enable the setting.
Setting the tolerance range alarm delay.
The current setting flashes Use the
arrow buttons
to enter the desired
time after which the range alarm shall be triggered. Entry range: 1 up to
300 minutes. Factory setting: 60 minutes.
Confirm the entry with the
OK button
.
Range alarm delay (m
Содержание ULTRA.Guard UF V 500
Страница 106: ...UF V E3 09 2018 Page 106 116 27 7 Dimensions UF V 500 E3 Dimensions in mm...
Страница 107: ...UF V E3 09 2018 Page 107 116 27 8 Dimensions UF V 700 E3 Dimensions in mm...
Страница 108: ...UF V E3 09 2018 Page 108 116 28 Certificates and declarations of conformity 28 1 EU Declaration of Conformity...
Страница 109: ...UF V E3 09 2018 Page 109 116...
Страница 110: ...UF V E3 09 2018 Page 110 116 29 Product registration...
Страница 117: ......