14.2 Information messages
Information messages provide information about settings made and the condition of the optional battery.
In Normal display a text message indicates the condition. The “Information” icon is lit.
Information message (example: CO
2
emergency cooling test run is
active)
CO2 cooling test acti
Information
messages overview:
Condition
Information message
Moment of information mes-
sage
Chamber with optional “battery-buffered
alarm system”:
very low battery voltage (1 V
to 11 V).
Chamber with options “battery-buffered
alarm system” and CO
2
emergency cooling:
Low battery voltage during CO
2
emergency
cooling on battery power.
Message on battery power and on line pow-
er.
“Low battery voltage”
immediately
Service setpoint is active
“Service setpoint active”
immediately
Emergency cooling test run is active
“CO2 cooling test active”
immediately
14.3 Activating / deactivating the audible alarm (alarm buzzer)
Path:
Normal display
Settings
Chamber
Audible alarm
Press the
OK button
to enable the setting.
Setting the audible alarm.
The current setting flashes. Use the
arrow buttons
to select between
ON and OFF.
Confirm the setting with the
OK button
.
Audible alarm
With the
Back button
you can go back to the “
Chamber
” submenu and, repeatedly pressing it, to
Normal
display
.
Содержание ULTRA.Guard UF V 500
Страница 106: ...UF V E3 09 2018 Page 106 116 27 7 Dimensions UF V 500 E3 Dimensions in mm...
Страница 107: ...UF V E3 09 2018 Page 107 116 27 8 Dimensions UF V 700 E3 Dimensions in mm...
Страница 108: ...UF V E3 09 2018 Page 108 116 28 Certificates and declarations of conformity 28 1 EU Declaration of Conformity...
Страница 109: ...UF V E3 09 2018 Page 109 116...
Страница 110: ...UF V E3 09 2018 Page 110 116 29 Product registration...
Страница 117: ......