UF V (E3) 09/2018
Page 38/116
5.1
Menu structure of the controller and access levels
Starting from Normal display, navigate between the menus with the
arrow buttons
.
With the
OK button
you enter the setting of further subordinate menu functions.
With the
Back button
you go back to the previous function and finally back to Normal display.
The available functions depend on the current dependent on the current
authorization
“User”, “Admin” or
“Service”, for which the entry of a password may be required, depending on the setting.
You can set passwords for different access levels:
•
User:
The password enables access to the standard operating functions. Factory setting is 00 00 (no
password assigned).
•
Admin:
The password enables access to advanced controller functions and settings. Factory setting is
00 01.
•
Service:
The password enables access to all controller functions (for BINDER Service only).
As soon as a password has been assigned, access to the respective functions is blocked and only availa-
ble after entering the correct password.
Menu
Required access level
Functions
Setpoints
“User”
•
Temperature set-point setting
•
CO
2
emergency cooling set-point setting (option
•
Setting the safety controller
•
Activating/deactivating the CO
2
emergency cooling (op-
tion) and the service setpoint, emergency cooling test (op-
tion)
Chamber info
Any user
•
Configuration display (setup information, controller hard-
ware and software, analog inputs)
•
Display of battery condition (“battery-buffered alarm sys-
tem” option)
•
Display of interface configuration (e.g. MAC address, IP
address)
Settings
“Admin”
•
General controller settings (date, time, menu language,
temperature unit, display brightness…)
•
Network settings
•
Setting the data logger storage interval
•
Setting the tolerance ranger limits and delay times for door
open and tolerance range alarm, setting the service set-
point
•
Access code assignment (“door access system” option)
•
Password changing for User and Admin
Service
“Service”
•
Configuration settings (only for BINDER Service)
•
Password changing for User and Admin
USB
Export: Any user
Import: “Admin”
•
Export of configuration, logger, and service data
•
Import of configuration data
Содержание ULTRA.Guard UF V 500
Страница 106: ...UF V E3 09 2018 Page 106 116 27 7 Dimensions UF V 500 E3 Dimensions in mm...
Страница 107: ...UF V E3 09 2018 Page 107 116 27 8 Dimensions UF V 700 E3 Dimensions in mm...
Страница 108: ...UF V E3 09 2018 Page 108 116 28 Certificates and declarations of conformity 28 1 EU Declaration of Conformity...
Страница 109: ...UF V E3 09 2018 Page 109 116...
Страница 110: ...UF V E3 09 2018 Page 110 116 29 Product registration...
Страница 117: ......