3.2.1 Equipotentiality between terminal and weighing platform
Make sure that there is equipotentiality between the metal parts of the
terminal and the weighing platform.
If there are doubts, connect the terminal and platform with an earth cable
of at least 6mm², using the ground screw at the rear of the terminal (see
point 1 Figure 3.1 on page 1-19).
The cables and accessories for this connection are not included in the
supply but are at the customer's charge.
3.2.2 Analogue cell connection
The pinout of connector JBIL when connected to platforms formed by
analogue cells is given below.
Key
NC = Stand-by. Do not connect
SIG + = + Signal
SIG - = - Signal
EX + = + Power supply
EX - = - Power supply
SENSE + = + SENSE signal
SENSE - = - SENSE signal
Figure 3.3 - JBIL connector pinout for connection to platforms
made with analogue cells (log0001.gif)
Venus
1-23