65
Electromagnetic Compatibility Information
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The IH 25/1 is intended for use in the electromagnetic environment specified below. The customer or the user of the IH 25/1 should assure
that it is used in such an environment.
Emissions
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The IH 25/1 uses RF energy only for its internal function. Therefore, its RF emis-
sions are very low and are not likely to cause any interference in nearby electro-
nic equipment.
RF emissions CISPR 11
Class B
The IH 25/1 is suitable for use in all establishments, including domestic esta-
blishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The IH 25/1 is intended for use in the electromagnetic environment specified below. The customer or the user of the IH 25/1 should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines ± 2 kV for power supply lines Mains power quality should be that of a typi-
cal commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) and neutral
± 1 kV line(s) and neutral
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5s
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5s
Mains power quality should be that of a typi-
cal commercial or hospital environment. If
the user of the IH 25/1 requires continued
operation during power mains interruptions,
it is recommended that the IH 25/1 be pow-
ered from an uninterruptible power supply
or a battery.
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8 3 A/m
3 A/m
Mains power quality should be that of a typi-
cal commercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level
Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity
The IH 25/1 is intended for use in the electromagnetic environment specified below. The customer or the user of the IH 25/1 should assure
that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
3 V
rms
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part
of the IH 25/1, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d
= 1,2
d
= 1,2
80 MHz to 800 MHz
d
= 2,3
800 MHz to 2,5 GHz
Where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagne-
tic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the IH 25/1 is used exceeds the applicable RF compliance level above, the IH 25/1 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the IH 25/1.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Содержание IH 25/1
Страница 48: ...48 50 51 1 Beurer 1 48 2 48 3 49 4 50 5 51 6 51 7 52 8 53 9 54 10 54 11 55 12 56 2 BF 2 I O SN IP21 12 5...
Страница 49: ...49 3 IH 25 1 Beurer Beurer...
Страница 50: ...50 30 4 1 2 3 4 5 6 3 1 2 6 4 5...
Страница 51: ...51 7 8 9 10 11 12 13 14 15 16 5 53 7 10 7 4 3 6 8 8 53 4 8 7 8 9 10 15 16 14 13 12 11...
Страница 52: ...52 8 1 8 8 8 2 4 2 8 3 8 11 12 13 14 6 2 4 5 2 8 6 6 53 7 10 1 5 1 2...
Страница 53: ...53 2 3 8 8 8 8 11 9 10 7 8 4...
Страница 54: ...54 10 4 9 WEEE Waste Electrical and Electronic Equipment 10 1 0 5 8 2 3 4...
Страница 67: ...67...
Страница 68: ...68 752 987_0114 Irrtum und nderungen vorbehalten...