35
If your smartphone has a protective cover, remove this to en-
sure that there is no interference during the transfer. Begin the
data transfer in the “beurer HealthManager”-app.
9. Cleaning and storing the device and cuff
• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries. The cuff line
should not be bent sharply.
10. Error messages/trouble-shooting
In case of faults, the _ message appears in the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
or
appears on the display)
• systolic or diastolic pressure was outside the measurement
range ( or
Lo
appears on the display)
• the cuff is fastened too tightly or loosely (
or
appears
on the display)
• the pump pressure is higher than 300 mmHg (
appears
on the display)
• pumping up takes longer than 160 seconds (
appears on
the display)
• there is a system or device error (
,
,
or
ap-
pears on the display)
• the data could not be sent to the PC (
appears in the
display).
• the data could not be sent via
Bluetooth
®
( ).
In such cases, repeat the measurement and/or the data trans-
fer. Ensure that the cuff line is correctly attached and that you
do not move or speak during the measurement.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section “Technical specifica-
tions”, the technical alarm will appear on the display indicating
either “
” or “
Lo
”. In such cases, you should seek medical as-
sistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.
Содержание BM 85
Страница 117: ...117 USB USB Beurer 1 1 117 2 118 3 122 4 124 5 125 6 128 7 130 8 132 9 133 10 133 11 134 12 134 13 136 14 137 15 137...
Страница 118: ...118 6 2 BF WEEE Waste Electrical and Electronic Equipment 21 PAP Storage Transport Operating...
Страница 119: ...119 SN CE 93 42 EEC 30...
Страница 120: ...120 3...
Страница 121: ...121 WEEE Waste Electrical and Electronic Equipment...
Страница 122: ...122 3 1 2 3 4 5 USB 6 7 M1 M2 8 9 10 11 1 2 3 4 10 9 5 11 6 7 8...
Страница 124: ...124 4 USB USB USB USB Bluetooth Pb Cd Hg Bluetooth Bluetooth M1 M2 5 M1 M2 M1 M2 M1 M2...
Страница 125: ...125 M1 M2 12 M1 M2 M1 M2 Bluetooth Bluetooth M1 M2 Bluetooth Bluetooth Bluetooth Bluetooth 5 2 3...
Страница 126: ...126 OK 22 42 3...
Страница 128: ...128 6 5 3 180 110 2 160 179 100 109 1 140 159 90 99...
Страница 129: ...129 130 139 85 89 120 129 80 84 120 80 WHO 1999 World Health Organization 5...
Страница 130: ...130 7 60 3 M1 M2 M1...
Страница 131: ...131 M2 Bluetooth M1 M2 1 M1 2 M2 A AM 7 5 00 9 00 M1 M2 PM 7 18 00 20 00...
Страница 137: ...137 14 22 42 163 946 071 62 USB 163 484 15...
Страница 157: ...157 Naprawy lub ca kowita wymiana w adnym wypadku nie prze d u aj okresu gwarancji Zastrzega si prawo do pomy ek i zmian...
Страница 158: ...158...
Страница 159: ...159...