6. Software Functions
61
6.6.2 Counter/Timer
If you select the measurement mode
Counter/Timer
, you will
see the following options:
Display for alpha beta contamination probe.
Display for dose rate probe.
Figure 6.8: Parameters for a Counter/Timer measurement,
for two different probes
Measuring Time
Measuring time of a cycle in seconds.
Cycles
The cycle parameters for all measurement modes are defined
in this menu:
One defines whether
the measurement is to be performed once (setting:
1)
the measurement is to be repeatedly constantly (setting
0)
a predefined number of cycles is to be performed (setting
N, where N can be between 2 and 9999).
Preset Acc. (%)
Once the accuracy of the measurement is less than the preset
accuracy, a measurement cycle is stopped even if the preset
measuring time has not yet been reached. If there are two
measuring channels (e.g. alpha and beta channel), the meas-
urement cycle is stopped as soon as both accuracies are below
the two preset accuracies. If you want to stop a measurement
via the preset measuring time, you have to enter a very small
value for the preset accuracy to make sure it will never be
reached, for example, 0.01%.
Auto Save
Enable (
X
) or disable (
) the automatic save function. Stored da-
ta can be viewed on the
Memory
menu either on the display or
Содержание LB 134
Страница 1: ...LB 134 Universal Monitor II Program version V 1 00 and higher ID No 62688BA2 Rev No 02 08 04 2016...
Страница 2: ......
Страница 6: ...LB 134 Content IV...
Страница 8: ...LB 134 Working with the User s Manual VI...
Страница 10: ...LB 134 Safety Instructions VIII...
Страница 34: ...LB 134 2 System Description 24 Figure 2 9 LB 134 with wall bracket...