![Berthold LB 134 Скачать руководство пользователя страница 100](http://html1.mh-extra.com/html/berthold/lb-134/lb-134_user-manual_2739260100.webp)
LB 134
7. Probes for the LB 134
90
All valves close automatically when the detector is lifted off the
bracket; this stops the gas supply and keeps the P10 gas in the
counting chamber.
The operation life after filling or flushing is at least 8 hours with
a loss in efficiency of less than 5 %. See also Figure 7.8 which
shows the efficiency of a counter tube filled with P10 gas over
the course of 3 days.
Efficacy of the P10 flow-through counter tube as a function
of time.
The following diagram shows the efficiency after a counter tube
filling with P10 gas, for alpha and also for beta sources, as a
function of the time. The average loss in efficiency for beta
sources is 0.1% / h. The average loss in efficiency for alpha
sources is 0.05 % / h.
The diagram illustrates that measurements over 8 hours are
easily possible with one counter tube filling with less than 5%
loss in efficiency.
Figure 7.8: Efficiency of a counter tube filled with P10 gas
as a function of time
Commissioning of the alpha-beta detector with P10 gas supply
1. First, unscrew the
valve opener
(counter-clockwise) from
one of the two valves of the P10 flow-through counter tube.
The valve opener, a screw with a borehole, ensures that no
over- or under-pressure builds up in the detector during
transport which may damage the detector foil. Keep the
Содержание LB 134
Страница 1: ...LB 134 Universal Monitor II Program version V 1 00 and higher ID No 62688BA2 Rev No 02 08 04 2016...
Страница 2: ......
Страница 6: ...LB 134 Content IV...
Страница 8: ...LB 134 Working with the User s Manual VI...
Страница 10: ...LB 134 Safety Instructions VIII...
Страница 34: ...LB 134 2 System Description 24 Figure 2 9 LB 134 with wall bracket...