![Berthold LB 134 Скачать руководство пользователя страница 145](http://html1.mh-extra.com/html/berthold/lb-134/lb-134_user-manual_2739260145.webp)
LB 134
11. Appendix
135
11.5.2 Transmission Protocol
Basic condition
The LB 134 can send and receive.
The response time of the LB 134 is at most 2 seconds.
The transmission pattern is as follows:
* TDV FDV PRC MBR TYP ITM (MSK Data)*CSM*CR
<
—Frame header———————>
<
—————————Frame—————————>
where
(..) optional
TDV
to device
0001 - 9999
FDV
from device
9001 - 9008
PRC
Process
00 - 03
MBR
Member
00 - 06
TYP
Type
00 - 03
ITM
Item
00 - 30
MSK
Mask on data 000000 - FFFFFF (not used)
Data
data
variable length
CSM
checksum
empty if no checksum
00 - FF
8 Bit checksum
CR
carriage return
(Minimum length = 20, maximum 255)
The stars "*" must always be included, even if CSM is empty.
The LB 134 answers, with TDV and FDV being exchanged in the frame.
The LB 134 does not respond to request, if
1. TDV does not match address
2. FDV is not between 9001 and 9008
3. The transmission pattern is incorrect
4. The checksum is not correct
Explanation of the transmission frame
1. TDV = logical address of LB 134 set by user program
2. FDV = logical address of host computer, set in the user program. Always 9001.
3. PRC = process
00 = system level
01 = read radiological data
03 = digital outputs on/off, reset
Содержание LB 134
Страница 1: ...LB 134 Universal Monitor II Program version V 1 00 and higher ID No 62688BA2 Rev No 02 08 04 2016...
Страница 2: ......
Страница 6: ...LB 134 Content IV...
Страница 8: ...LB 134 Working with the User s Manual VI...
Страница 10: ...LB 134 Safety Instructions VIII...
Страница 34: ...LB 134 2 System Description 24 Figure 2 9 LB 134 with wall bracket...