![Berthold LB 134 Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/berthold/lb-134/lb-134_user-manual_2739260057.webp)
LB 134
5. Contamination Measurements
47
In the
Display Mode
menu you can change the display lay-
out.
Then select the nuclide to be measured:
-
If two measuring channels are displayed, press the
Nuclides
key first. Then the function of the softkey
changes. Using the keys
and
you can now set the
and the
nuclides separately.
Figure 4.15: Nuclide selection with simultaneous
/
measurement
In the selected measurement mode (Bq/cm²), the detected
count rates (cps) are converted to area activities (Bq/cm²) using
a calibration factor entered in the device. The previously saved
background rate is automatically subtracted, so that the net ar-
ea activities are displayed here.
Press the
Info
softkey to view further information on the ongo-
ing measurement. The
Graph
function allows you to present
the change of the measured value over time graphically. The
graph runs from right to left and you always see the values of
the last 120 seconds. With
Delete
the graph can be reset and it
starts all over again. In the counter/timer mode we get a new
pixel in the graph for each elapsed cycle.
Figure 4.16: Change in the
-values over time displayed as a
graph
Содержание LB 134
Страница 1: ...LB 134 Universal Monitor II Program version V 1 00 and higher ID No 62688BA2 Rev No 02 08 04 2016...
Страница 2: ......
Страница 6: ...LB 134 Content IV...
Страница 8: ...LB 134 Working with the User s Manual VI...
Страница 10: ...LB 134 Safety Instructions VIII...
Страница 34: ...LB 134 2 System Description 24 Figure 2 9 LB 134 with wall bracket...