8 Error Codes and Alarms
LB 571 BTwave
42
68847BA2 Rev.00, 10/2020
8.3
Alarm generation
8.4
Enabling alarms
Error codes causing an alarm can be defined using
device_parameters #70
(Alarm_mask)
. By setting corresponding bit high in mask defines that error is taken
into account for alarm generation. Multiple errors can be defined by setting mul-
tiple bits high.
Alarm mask examples
Alarm mask
value (hex)
Alarm mask value (dec)
Error code(s) to cause alarm
0x0001
1
SENSOR_RESONANCE_LEVEL_ERR
0x0002
2
SENSOR_TEMP_MATERIAL_ERR
0x0004
4
SENSOR_TEMP_CERAME_ERR
0x0008
8
SENSOR_ANALOG_OUT_1_ERR
0x0010
16
SENSOR_ANALOG_OUT_2_ERR
0x0020
32
SENSOR_SD_CARD_WRITE_ERR
0x0040
64
SENSOR_PLL_CONFIG_ERR
0x0080
128
SENSOR_RTC_ERR
0x0100
256
SENSOR_WD_RESET_ERR
0x0200
512
SENSOR_HEATING_CTRL_ERR
0x0400
1024
SENSOR_SD_CARD_INIT_ERR
0x0087
(1+2+4+128) = 135
SENSOR_RESONANCE_LEVEL_ERR
SENSOR_TEMP_MATERIAL_ERR
SENSOR_TEMP_CERAME_ERR
SENSOR_RTC_ERR
0x07FF
(1+2+4+8+16+32+64+128+256
+512+1024)=2047
All errors generate alarm
Delay prior alarm is triggered from occurrence of bitwise OR of errors and can be
defined from
device_params #71 (Alarm_delay)
.
When alarm conditions disappears, alarm is switched off after the same
de-
vice_params #71 (Alarm_delay)
. Watchdog error needs power down of device for
getting cleared.
If device alarm output is triggered, connect to the device using the UI and check
active errors. Below are some checkpoints in case of errors.
SENSOR_RESONANCE_LEVEL_ERR
Check parameters and that measurement signal is ok during material measurement
(above -50 dB). If not, empty resonator surface and check that signal is ok and ap-
prox. -10 dB (see chapter 10.1). Change device.
SENSOR_TEMP_MATERIAL_ERR
Check that material temperature reading is ok, otherwise try changing tempera-
ture sensor. Change device.
SENSOR_TEMP_CERAME_ERR
Check that ceramic temperature reading is ok, otherwise try powering device off,
if it does not help, change device.
SENSOR_ANALOG_OUT_x_ERR
Check that connections and parameters are ok. If connection is not needed, disable
output channel by setting AOx_I_min to -1.
Содержание BTwave LB 571-02
Страница 1: ...BTwave LB 571 02 Operating Manual 68847BA2 Rev No 00 01 2020 ...
Страница 2: ......
Страница 4: ......
Страница 24: ...5 Commissioning LB 571 BTwave 20 68847BA2 Rev 00 10 2020 ...
Страница 44: ...7 Diagnostics LB 571 BTwave 40 68847BA2 Rev 00 10 2020 ...
Страница 48: ...8 Error Codes and Alarms LB 571 BTwave 44 68847BA2 Rev 00 10 2020 ...
Страница 50: ...9 Technical Information LB 571 BTwave 46 68847BA2 Rev 00 10 2020 ...
Страница 51: ...LB 571 BTwave 10 Declaration of Conformity 68847BA2 Rev 00 10 2020 47 10Declaration of Conformity ...