LB 571 BTwave
8 Error Codes and Alarms
68847BA2 Rev.00, 10/2020
43
SENSOR_SD_CARD_XXX_ERR
Check that SD-card is inserted. Change SD-card.
SENSOR_PLL_CONFIG_ERR
Check that parameters are ok. Try powering device off, if it does not help, change
device.
SENSOR_RTC_ERR
Set device time to PC time. See chapter 6.3.
SENSOR_WD_RESET_ERR
Check that parameters are ok. Try powering device off, if it does not help, change
device.
SENSOR_HEATING_CTRL_ERR
Check that ceramic temperature reading is ok. Check heater connections (see chap-
ter 3.1). Change device.