Configuring and Operating the Driving Unit
en
WARNING
8.5
Configuring the System
Only configure blood pumps and cannulae that are being used. Deviat-
ing settings could lead to the patient receiving insufficient support.
If there is no way of adjusting insufficient driving unit settings, take al-
ternative corrective actions immediately. Possible corrective actions
could be supplying the patient with the manual pump or switching to an-
other driving system.
Audible alarm signals are only emitted during the start test and when
the blood pumps are in automatic operation. An audible alarm is not
emitted in single-step mode or when one or both blood pumps are
stopped. The driving unit must thus be monitored in such cases.
Otherwise, alarms could be noticed too late or not at all.
Check the blood pumps and cannulae for full de-airing before
A
UTO
op-
eration. Perform individual steps to de-air and check the system.
8.5.1
Configuration Options - Overview
After switching on the system, the driving unit can be configured. Three options are
available.
Configuration
Prerequisite
M
ANUAL
—
F
ROM
F
ILE
A suitable configuration file is saved on the panel
PC or the USB stick.
Commissioning from a file makes sense if very
few parameters need to be changed. In particu-
lar, the BVAD, LVAD or RVAD operating mode
must be the same.
R
EPLACEMENT
The configuration file for the primary driving unit
is saved on the panel PC or the USB stick.
Tab. 8-3
Configuration - options
Содержание EXCOR Adult
Страница 2: ...2 1015002x00 en ...
Страница 12: ...12 1015002x00 Table of Contents en ...
Страница 14: ...14 1015002x00 Contact en ...
Страница 66: ...66 1015002x00 Driving Unit en ...
Страница 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Страница 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Страница 126: ...126 1015002x00 Replacing the Driving System en ...
Страница 128: ...128 1015002x00 Weaning and Explantation en ...
Страница 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Страница 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Страница 154: ...154 1015002x00 Manual Pump en ...
Страница 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Страница 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Страница 194: ...194 1015002x00 Technical Specifications en ...
Страница 198: ...198 1015002x00 Overview Parameters en ...
Страница 206: ...206 1015002x00 EMC en ...
Страница 208: ...208 1015002x00 Patient ID Labels en ...
Страница 210: ...Abbrevations 210 1015002x00 en ...
Страница 212: ...212 1015002x00 en ...