1015002x00
47
Driving Unit
en
5.2.3.7 Guiding the Driving Tube and the Flow Sensor Cable in the Cable Wrap
1.
Draw the cable wrap up from the
blood pump toward the driving
unit.
2.
Align the driving tube and sen-
sor cable.
3.
Insert the driving tube and the
sensor cable into the wrapping
tool.
4.
Attach the cable wrap to the
wrapping tool
5.
Slowly guide the wrapping tool
along the cable wrap with one
hand while grasping the cable
wrap with the other. Avoid creat-
ing tensile force on the blood
pump or the cannulae.
6.
At the end, guide the driving
tube and sensor cable out of the
wrapping tool. The sensor cable
will need to form a small loop by
the driving unit. The loop makes
plug insertions easier.
7.
Insert the flow sensor into the
corresponding connector. Make
sure to keep left and right in
mind.
Содержание EXCOR Adult
Страница 2: ...2 1015002x00 en ...
Страница 12: ...12 1015002x00 Table of Contents en ...
Страница 14: ...14 1015002x00 Contact en ...
Страница 66: ...66 1015002x00 Driving Unit en ...
Страница 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Страница 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Страница 126: ...126 1015002x00 Replacing the Driving System en ...
Страница 128: ...128 1015002x00 Weaning and Explantation en ...
Страница 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Страница 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Страница 154: ...154 1015002x00 Manual Pump en ...
Страница 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Страница 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Страница 194: ...194 1015002x00 Technical Specifications en ...
Страница 198: ...198 1015002x00 Overview Parameters en ...
Страница 206: ...206 1015002x00 EMC en ...
Страница 208: ...208 1015002x00 Patient ID Labels en ...
Страница 210: ...Abbrevations 210 1015002x00 en ...
Страница 212: ...212 1015002x00 en ...