56
1015002x00
Driving Unit
en
5.7.1.9 Driving Unit Operating Time in Battery Operation
The operating time of the driving unit depends primarily on the operating mode, the
size of the blood pump, the set rate, and the ambient temperature. The cannulae in
use, the systolic duration, and the patient’s pre- and afterloads all affect operating time
as well.
The following values are guidelines for biventricular operation of 10 mL blood pumps
or an 80 mL/60 mL combination.
BVAD 10 mL blood pumps
Ambient
temperature
Battery oper-
ating time
Operating
time
Emergency
battery
Cannulae 6 mm / 5 mm
Rates 90 bpm / 80 bpm
Systole 30 % / 40 %
Preload 10 mmHg
Afterload 100 mmHg
0 °C
min. 8 h
at least 30
min
25 °C
min. 9 h
40 °C
min. 8:30 h
Operating times are based on new, 100% charged batteries. The operating time
of the emergency battery is measured from the “P7 Emergency battery dis-
charged” notification.
Tab. 5-4
Driving unit operating time in battery operation or emergency battery operation, 10 mL
BVAD 80 mL / 60 mL
Blood pumps
Ambient
temperature
Battery oper-
ating time
Operating
time
Emergency
battery
Cannulae 12 mm / 12 mm
Rates 90 bpm / 80 bpm
Systole 50 %
Preload 10 mmHg
Afterload 100 mmHg
0 °C
min. 5 h
at least 30
min
25 °C
min. 5:15 h
40 °C
min. 5:15 h
Operating times are based on new, 100% charged batteries. The operating time
of the emergency battery is measured from the “P7 Emergency battery dis-
charged” notification.
Tab. 5-5
Driving unit operating time in battery operation or emergency battery operation, 80 mL / 60 mL
Содержание EXCOR Adult
Страница 2: ...2 1015002x00 en ...
Страница 12: ...12 1015002x00 Table of Contents en ...
Страница 14: ...14 1015002x00 Contact en ...
Страница 66: ...66 1015002x00 Driving Unit en ...
Страница 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Страница 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Страница 126: ...126 1015002x00 Replacing the Driving System en ...
Страница 128: ...128 1015002x00 Weaning and Explantation en ...
Страница 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Страница 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Страница 154: ...154 1015002x00 Manual Pump en ...
Страница 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Страница 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Страница 194: ...194 1015002x00 Technical Specifications en ...
Страница 198: ...198 1015002x00 Overview Parameters en ...
Страница 206: ...206 1015002x00 EMC en ...
Страница 208: ...208 1015002x00 Patient ID Labels en ...
Страница 210: ...Abbrevations 210 1015002x00 en ...
Страница 212: ...212 1015002x00 en ...