Case Expansion Monitor Maintenance
1 8 1 Alarm Setpoint Adjust
84418-01
I
The 3300 Case Expansion Monitor has alarm setpoints for the differential values only. The
alarm setpoints are adjusted on the left side of the display. To adjust the differential alarm
setpoints, use the following procedure.
1. Open front panel.
2. Set switch AA (Adjust Alarms) to the
left (ON). T h e bar graph will start
flashing.
3. To adjust ALERT or DANGER set
points, press a n d hold ALERT o r
DANGER switch on the .front panel.
4. Use the (r) or
(t)
switches on the
System Monitor to adjust the setpoint
level up or down.
5. Reset AA to the right (OFF); close
front panel.
13