BENNING MA 3 measuring adapter

 

 

 

5.  Connecting the BENNING MA 3 

 

 

The BENNING MA 3 measuring adapter is 
equipped with a PTC thermistor as short-circuit 
protection. 

As the PTC thermistor limits the maximum cur-
rent, it is not possible to carry out functional 
tests according to VDE 0701-0702 with the 
BENNING MA 3 measuring adapter. 

Use the BENNING MA 3 measuring adapter 
only for the intended tests described in this 
manual.  

 

 

The manufacturers’ specifications or normative 
restrictions with regard to all tests involved must 
be adhered to. 

To do this, thoroughly read the notes stated in 
the corresponding standard, in the manufactur-
ers’ specifications of the test sample and in the 
operating manual of the appliance tester. 

 

Before you can start testing, you have to connect the  

BENNING MA 3

  measuring adapter to your appliance 

tester. 

To do this, proceed as follows: 

  Make sure that the appliance tester you are using is 

ready for operation. 

  Connect the shock-proof plug of the 

BENNING MA 3

 

measuring adapter to the test socket of your appli-
ance tester. 

 

Connecting the BENNING MA 3 

Connecting the BENNING MA 3 to the appliance tester 

 

 

 

Connection diagram 

 
 

6.  Testing with the BENNING MA 3 

6.1. Testing of three-phase devices 

 

It is possible to test single-phase devices in the 
same way. 

 

6.1.1.  RPE – Protective conductor resistance 

Before testing 

 

 

Disconnect the test sample from the mains 
supply! 

 

 

The maximum testing current for RPE meas-
urement is 10 A

 

 

The rotary switch of the 
BENNING MA 3 can be 
used for RPE measure-
ment in any locking posi-
tion. 

 

 

Before starting the test, determine the internal 

RPE

  re-

sistance of the measuring adapter. 

Internal RPE resistance of the measuring adapter 

RPE – BENNING MA 3 

 

  Start the 

RPE

 measurement on your appliance 

tester. 

  Contact the test probe of the appliance tester with 

the PE contact 

(pos. 10, chapter 4 “Operating elements”) 

of the corresponding CEE 

socket

  

(see figure below “

RPE-BENNING MA 3

”). 

  Write down the measured 

RPE

 value. 

  Follow the test procedure 

Class I, RPE

 for “

porta-

ble test samples

”. 

 

RPE – BENNING MA 3 (e. g. CEE 5-pin, 32 A) 

Setup for measuring the internal RPE resistance of the MA 3 

 

 

 

Connection diagram 

 

 

Detailed view 

 
 

 

Содержание 044159

Страница 1: ...BENNING MA 3 CEE Messadapter 5 3 polig Operating manual Translated version English 5076 en Version 1 00 19 February 2018...

Страница 2: ......

Страница 3: ...NG MA 3 3 6 1 Testing of three phase devices 3 6 1 1 RPE Protective conductor resistance 3 6 1 2 RInsu Insulating resistance 4 6 1 3 Alternative leakage current measurements 6 6 2 Testing of extension...

Страница 4: ...more than one test sample or one extension cable at a time In case of proper use according to the specifications the safety of both the user and the device is ensured Please read the operating manual...

Страница 5: ...V L N PE 6h 3 CEE connector 32 A 50 Hz to 60 Hz 380 V to 415 V 3L N PE 6h 4 Status LED indications CEE connector 3 pin 5 pin 5 Rotary switch for RInsu test alternative leakage current test functional...

Страница 6: ...NNING MA 3 measuring adapter to the test socket of your appli ance tester Connecting the BENNING MA 3 Connecting the BENNING MA 3 to the appliance tester Connection diagram 6 Testing with the BENNING...

Страница 7: ...est sample from the mains supply Switch the rotary switch of the BENNING MA 3 to the locking position L1 L2 L3 Test procedure VDE 0701 0702 Class I RInsu Portable test sample Connect the CEE connector...

Страница 8: ...am Circuit diagram Class I RInsu secondary to PE Portable test sample Connection diagram Circuit diagram Class I RInsu LN to secondary Portable test sample Connection diagram Circuit diagram Class I R...

Страница 9: ...d appliance tester IPE protective conductor current ICont contact current ILeak device leakage current IPLeak patient leakage current Before testing Disconnect the test sample from the mains supply Sw...

Страница 10: ...on class II that raise concerns as to an RInsu measure ment Moreover it is checked to prove the proper insula tion capacity under mains voltage This also applies to de vices of protection class I with...

Страница 11: ...t sample Application Class I IPLeak patient leakage current Portable test sample Connection diagram Circuit diagram Test procedure VDE 0751 1 Class II IPLeak patient leakage current Portable test samp...

Страница 12: ...CEE connections of the BENNING MA 3 for future measurements and add them to the measuring adapter RPE BENNING MA 3 e g CEE 5 pin 32 A Setup for measuring the internal RPE resistance of the MA 3 Conne...

Страница 13: ...ductor is affected Application Class I RInsu LN to PE Cable test Connection diagram Circuit diagram 6 2 3 Functional test and phase sequence test of cables Before testing The functional test and phase...

Страница 14: ...C to 60 0 C Max relative humidity linearly decreasing non condensing 80 0 at 30 0 C 60 0 at 40 0 C Protection category IP 40 with the case being open IP 67 with the case being closed Dimensions W x D...

Страница 15: ...areparts benning de General service requests Phone 49 2871 93 556 E mail servicerequests benning de Returns management Phone 49 2871 93 554 E mail returns benning de Training management Phone 49 2871...

Страница 16: ...to the technical status at the time of printing Subject to technical changes No liability for printing errors Support helpdesk BENNING helpdesk team Phone 49 2871 93 555 Fax 49 2871 93 417 E mail info...

Отзывы: