BENNING MA 3 measuring adapter
7
Test procedure VDE 0701-0702
Class I; ICont – contact current
Portable test sample
Connect the CEE connector of the test sample to the
CEE
socket
of the
BENNING MA 3
.
Start the
ICont
measurement on your appliance
tester.
Contact the test probe(s) of the appliance tester with
all necessary measuring points
As a result, you will obtain the measured
ICont
value of
the test sample.
Application
Class I; ICont – contact current
Portable test sample
Connection diagram
Circuit diagram
The contact current is determined for devices of
protec-
tion class II
that raise concerns as to an
RInsu
measure-
ment. Moreover, it is checked to prove the proper insula-
tion capacity under mains voltage. This also applies to de-
vices of
protection class I
with accessible conductive
parts which are not connected to the protective conductor.
Test procedure VDE 0751-1
Class I; ILeak – device leakage current
Portable test sample
Connect the CEE connector of the test sample to the
CEE
socket
of the
BENNING MA 3
.
Start the
ILeak
measurement on your appliance
tester.
Contact the test probe(s) of the appliance tester with
all necessary measuring points
As a result, you will obtain the measured
ILeak
value of
the test sample.
Application
Class I; ILeak – device leakage current
Portable test sample
Connection diagram
Circuit diagram