BENNING MA 3 measuring adapter
6
Class II; RInsu; LN to bodies
Portable test sample
Connection diagram
Circuit diagram
Class II; RInsu; LN to secondary
Portable test sample
Connection diagram
Circuit diagram
6.1.3. Alternative leakage current measurements
The
BENNING MA 3
can be used for measuring the fol-
lowing leakage or fault currents by means of the alterna-
tive leakage current measurement method, provided that
the test is supported by the connected appliance tester:
•
IPE
– protective conductor current
•
ICont
– contact current
•
ILeak
– device leakage current
•
IPLeak
– patient leakage current
Before testing
Disconnect the test sample from the mains
supply.
Switch the rotary switch of
the BENNING MA 3 to the
locking position “L1-L2-L3”.
Test procedure VDE 0701-0702
Class I; IPE
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
BENNING MA 3
.
Start the
IPE
measurement on your appliance tester.
As a result, you will obtain the measured
IPE
value of the
test sample.
Application
Class I; IPE – protective conductor current
Portable test sample
Connection diagram
Circuit diagram