7.2 Test steps
A classified test sequence with the UNIMET
®
300ST comprises a number of individual test steps. The
test steps are sorted on the basis of measurement numbers.
Meas
No.
Name of measurement
Short text
Unit
Standard
3
PE conductor resistance, permanently
attached cable
R-PE
Ohm
VDE 0701-0702
+ IEC 62353
4
Protective earth resistance, installed
equipment
R-PE
Ohm
VDE 0701-0702
+ DIN EN 62353
5
Insulation resistance, Protection Class II
R-ISO CII
MOhm
IEC 62353
6
Insulation resistance, Protection Class I
R-ISO CI
MOhm
IEC 62353
106
Insulation resistance, Protection Class I
R-ISO CI
MOhm
VDE 0701-0702
107
Insulation resistance, Protection Class II
R-ISO CII
MOhm
VDE 0701-0702
120
Insulation resistance, cable
R-ISO
MOhm
VDE 0701-0702
44
Equipment leakage current Protection
Class I - alternative measurement
I-ELC-A CI
mA
IEC 62353
46
Equipment leakage current -
Protection Class II - alternative
measurement
I-ELC-A CII
mA
IEC 62353
109
Equipment leakage current Protection
Class 1 -alternative measurement
I-ELC-A CI
mA
VDE 0701-0702
110
Equipment leakage current Protection
Class II -alternative measurement
I-ELC-A CII
mA
VDE 0701-0702
80
Load current measurement
A
VDE 0701-0702
+ DIN EN 62353
81
Operating voltage
V
VDE 0701-0702
+ DIN EN 62353
82
Power consumption
kVA
VDE 0701-0702
+ DIN EN 62353
111
PE conductor current - Direct
measurement
I-PE Direct
mA
VDE 0701-0702
112
PE conductor current - Direct
measurement - Phase reversed
I-PE Direct Ph-r
mA
VDE 0701-0702
113
PE conductor current - Differential
measurement
I-PE Differential
mA
VDE 0701-0702
114
PE conductor current - Differential
measurement - Phase reversed
I-PE Differential Ph-r
mA
VDE 0701-0702