Operation and configuration
15
4.2.2
Operation via the keyboard
An external keyboard makes it easy to enter the names and ID numbers of the test engineers. You
can also operate the UNIMET entirely via the keyboard. The Cursor block "
", "
", "
" "ENTER"
and "ESC" keys have the same function as the buttons on the front of the UNIMET.
Note when entering text:
The text entry cannot be longer than 20 characters. The UNIMET will truncate longer entries to
20 characters.
You can enter all characters which can be selected directly on UNIMET.
Upper and lower case is available but not umlauts.
The numeric keypad can only be used to enter digits. The LED "Num-Lock" therefore lights up
constantly.
When entering text, the following keys can be used to edit the text:
"
", "
" (Cursor block), "
" (Backspace), Del, Home.
Entries via the keyboard can also be combined with texts read in from the barcode scanner and
barcode reading wand.
Save the entry after entering the last character by
– by holding down the Enter button "
" on the UNIMET for approx. 3 seconds
– or by pressing the "
" ENTER key on the keyboard once.
4.2.3
Reading in with the barcode reader
Refer to the operating instructions for the barcode reading wand or barcode scanner.
The AR100 barcode reading wand (see “Ordering data” auf Seite 57.) reads more reliably by sweep-
ing it quickly over the barcode. Also follow the instructions for the barcode reading wand. Barcodes
can be read both forwards and backwards.
The data string which is read in can be up to 20 characters long. The UNIMET will truncate longer en-
tries to 20 characters.
When reading in the names and ID numbers of new test engineers, the input screen will remain
open. This allows you to check the entry and change or replace it if necessary. .
You can delete the entry and go to the previous menu by holding down
– the "ESC" button on the UNIMET for approx. 3 seconds
– or by pressing the "ESC" key on the external keyboard once.
Save the entry by holding down
– the Enter button "
" on the UNIMET for approx. 3 seconds
– or by pressing the "
" Enter key on the external keyboard once.
You can use the barcode scanner or barcode reading wand to call up tests from the device catalogue
again using the ID number. In this case, after successfully reading in the barcode, it will immediately
switch to the next screen of the operating process.