Testing and measuring
40
On completion of the test sequence, the functional test is carried out. Here, the DUT is put into
operation via the test socket on the test system.
5.1.3 Evaluating the test result
If "Device type query" in the device settings is enabled, this is entered before the device ID.
Message appears when
at least one test step
has not been passed.
passed
not passed
6. Evaluate test result
5. Start function test and
then evaluate result
***************************
TEST PASSED!
***************************
Next:
All
test steps
passed
?
yes
no
Volt. at sock.outlet
1. Save
2. Print
3. Save and print
4. View result
5. Cancel test
FUNCTIONAL TEST
1. Passed
2. Failed
3. Start Funct. test
***************************
TEST NOT PASSED!
***************************
Next:
ESC
Save,
Print,
Save and print
ESC
short = Edit previous characters
ESC
long = Exit menu without making change
short = Accept characters
long = Accept result, carry out
selected action and exit menu
ENTER DEVICE-ID:
Char.: A-Z,0-9,-, /
Exit:ESC Next:
TEST-RESULTS
Total result:
PASSED
Select:
,
Cont.:
View
result
Discard
test
,
= select next or previous test step
= back to previous menu
1. Save
2. Print
3. Save and print
4. View result
5. Cancel test