COMTRAXX® CP305 - Control Panel
2
Intended use
Remote alarm indicator COMTRAXX
®
CP305 is used for visual and audible signalling of
operating and alarm messages from the Bender systems MEDICS
®
, ATICS
®
, EDS and RCMS. In
MEDICS
®
monitoring systems, the CP305 meets the requirements of standard DIN VDE 0100-710
with regard to test functions for IT system monitoring and alarms from changeover devices. IT
system monitoring equipment are tested via the test button and the devices to be tested.
Important display functions
•
Normal operation indicator
•
Insulation fault
•
Overload
•
Overtemperature
•
Interruption of the system conductor or PE conductor connection of the ISOMETER
®
•
Supply line failure
•
Power supply fault conditions and changeover device faults
•
Hazard current
•
Device failure
•
Test results
•
Measured values
The connection between the CP350 and the transfer switching and monitoring modules is
implemented with bus technology. During normal operation, the CP305 indicates the readiness
for operation of the system.
Version CP305-IO features 12 digital inputs, allowing messages from other technical equipment
to be recorded and displayed on the CP305, such as from medical gases or battery-supported
safety power supply systems (BSV systems). Two relay contacts enable messages to be forwarded
to the higher-level building management system (BMS).
CP305 devices are used in
•
Healthcare facilities
•
Industrial and office buildings
•
Public buildings
Please heed the limits of the area of application indicated in the technical specifications.
Operator control and adjustment of settings is partly done directly on the device or in a web
browser.
Any other use than that described in this manual is regarded as improper.
Intended use includes:
•
Equipment-specific settings compliant with local equipment and operating conditions
•
Observation of all information in the operating manual
•
Compliance with test intervals
CP305_D00425_00_M_XXEN_09.2022