5.2
Regular inspection and cleaning
To ensure splicing quality, regular inspection and cleaning are required.
5.2.1
Object lens cleaning
Contamination on the object lens surface disturbs the identification of fiber core location
and consequently causes high splice loss. Object lenses should be kept clean at all times.
If dust accumulates for a prolonged period, it may be difficult to remove. Clean the lens
frequently as follows.
i.
Turn the power off before cleaning the object lens.
ii.
Separate the electrodes.
iii.
Use a soft cotton swab moistened with alcohol to clean in a circular motion from
the center as shown in the figure below. Dry the remaining alcohol using a clean,
dry, lint-free cotton swab.
iv.
The surface of the object lens should be clean without any lines or stains.
v.
Reassemble the electrodes.
vi.
Turn the power on; check for any lines or stains on the monitor; conduct a
self- diagnosis.
Содержание FX Fusion Splicer
Страница 12: ...PX106545 EN Rev A 10 3 2 Component name Sleeve heater Wind cover Monitor Battery...
Страница 13: ...PX106545 EN Rev A 11 Heater Heater cover Instructions for Use DC in charge in USB...
Страница 27: ...PX106545 EN Rev A 25 Deleting splice mode i Select a mode to be deleted ii Delete it by pressing...
Страница 29: ...PX106545 EN Rev A 27 Deleting heater mode i Select a mode to be deleted ii Delete it by pressing...
Страница 40: ...PX106545 EN Rev A 38 6 3 2 Deletion of splice result Data and images can be deleted in a single step...