C-3
CIIL COMMANDS
EXPLANATION
-
ALL SET, SRX,SRN
commands are part of the setup command above.
SET VOLT
<value>
Sets output voltage to given value, in volts rms. The following values are
permitted:
0 < value < 135, or 270, dependent on range selected and type of unit.
If set volt (value) not received, then use
SRN VOLT
(value), or
SRX
VOLT
(value). If none of these 3 are sent then generate error message.
SET FREQ
<value>
Optional. Sets frequency of output voltage to given value, in Hz. The
following values are permitted: 45 < value < 500. Default is 45 Hz only if
SET FREQ, SRN FREQ
, or
SRX FREQ
are not received. Otherwise the
values shall be used in the order stated.
SRX VOLT
<value>
Optional. Sets maximum limit for
SET VOLT
command. If
SET VOLT
value is greater than
SRX VOLT
value in a command string, an error
message is generated and the setup command ignored. The following
values are permitted:
0 < value < 135, or 270, dependent on range selected and type of unit.
If a value for
SRX VOLT
is not specified, the required maximum value for
SET VOLT
default to 135, or 270, dependant on range selected and type
of unit.
SRN VOLT
<value>
Optional. Sets minimum value limit for
SET VOLT
command. If
SET VOLT
value is less than
SRN VOLT
value in a command string, an error
message is generated and a setup command ignored. The following
values are permitted:
0 < value < 135, or 270, dependent on range selected and type of unit.
The
SET VOLT
default is 0 for the minimum value if a value is not
specified for
SRN VOLT
.
SRX FREQ
<value>
Optional. Sets maximum limit for
SET FREQ
command. If the
SET
FREQ
value in a command string is grater than the maximum limit, an
error message is generated and the setup command ignored. Legal
values are:
45 < value < 500.
If a value for
SRX FREQ
is not specified,
SRX FREQ
defaults to 500.
SRN FREQ
<value>
Optional. Assigns minimum value to
SET FREQ
command. If the
SET
FREQ
value in a command string is less than the minimum limit, an
error message is generated and the setup command ignored. Legal
values are:
45 < value < 500.
If a value for
SRN FREQ
is not specified,
SRN FREQ
defaults to 45.
SET VLT0
Set LO voltage range.
SET VLT1
Set HI voltage range.
Содержание BL10000 Series
Страница 9: ...3 THIS PAGE BLANK FOR ADDENDUMS ALL MANUAL ADDENDUMS WILL FOLLOW THIS PAGE ...
Страница 17: ...11 Figure 3 1 BL10000 AC Source Control Chassis Front and Rear Panel Views ...
Страница 49: ...6 1 SECTION 6 DRAWINGS ...
Страница 50: ......
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......