xi
Introduction
SPECIFICATIONS
Only values with tolerances or limits are guaranteed data. Values
without tolerances are informative data, without performance
guarantee.
CONTROL FEATURES
Speed
Set speed . . . . . . . . . . . . . . . . . . . . . 1000 rpm to 80 000 rpm in increments of 100 rpm
Speed control . . . . . . . . . . . . . . . . . . actual rotor speed will be ±50 rpm of the set speed (above 1000 rpm)
Speed display . . . . . . . . . . . . . . . . . . indicates rotor speed in increments of 10 rpm at speeds below 1000 rpm,
and increments of 100 rpm at speed above 1000 rpm
Time
Set time . . . . . . . . . . . . . . . . . . . . . . up to 99 hours and 59 minutes; HOLD for runs of unspecified length
Time display. . . . . . . . . . . . . . . . . . . indicates time remaining in timed runs, time elapsed in HOLD runs, or
estimated time remaining in
ω
2
t
runs
Rotor Temperature
Set temperature . . . . . . . . . . . . . . . . 0 to 40°C in increments of 1°C
Temperature control . . . . . . . . . . . . . ±1°C of set temperature
Temperature display. . . . . . . . . . . . . indicates rotor temperature in increments of 0.1°C
Temperature stability . . . . . . . . . . . . within 0.3°C of set after equilibrium is reached
STOP
VACUUM–
µ
SPEED–RPM
TIME–HR:MIN
TEMP–˚C
POWER OFF
X 10,000
SAVE
PROG
HOLD
w
2
t
DECEL
ACCEL
VACUUM
START
STOP
TIME
TEMP
CE
0
1
2
3
4
5
6
7
8
9
SPEED
ENTER/RECAL
L
Содержание Optima LE-80K