Required Wire Connections
To ensure safety, the instrument should be wired to a remote emer-
gency switch (preferably outside the room where the centrifuge is
housed, or adjacent to the exit from that room), in order to disconnect
the instrument from the main power source in case of a malfunction.
Space
Locate the Optima LE-80K in a clean, safe, uncluttered environment.
Be sure to provide a 5.1-cm (2-in.) clearance on each side of the
instrument, as the feet extend about 5.1 cm (2 in.) beyond the instru-
ment. (The pads under each foot have been designed to prevent
possible rotation of the instrument in the event of a rotor mishap.)
A 15.2-cm (6-in.) clearance is required at the rear of the instrument
for servicing the control panel and to ensure sufficient air ventilation
during operation.
The instrument will operate within specifications in a laboratory with
ambient temperatures ranging from 15 to 40°C.
WARNING
!
Do not place the instrument near areas
containing flammable reagents or
combustible fluids.
Wire Insulation
Color
Terminal
Symbol
Harmonized
North American
Green/Yellow
Earth ground
Light Blue
Neutral
N
L
Brown
Live or Line
L
L