4-4
Troubleshooting and Maintenance
Figure 4-2. Depressing the Latch to Loosen the Front Panel
4. Vent the vacuum chamber by slowly turning the small setscrew on
the vacuum solenoid valve to the right (clockwise) until you hear
the air released into the chamber. If you hear a whining noise,
close the valve and wait, because the rotor is still spinning.
5. About 10 seconds after the hissing noise stops, turn the screw to
the left (counterclockwise) until it stops.
6. Push down on the interlock pin (Figure 4-3) with your finger while
opening the chamber door with your free hand. (If the rotor is still
spinning, close the door and wait.) Be very careful not to touch
any wires or circuitry.
ENTER/RECALL
7
8
9
4
5
6
1
2
3
0
CE
VACUUM
SPEED
TIME
TEMP
START
STOP
Front Panel Tabs
Front Panel Access Port
Top Panel Tabs