2-10
Run Preparation
To use the delayed start program function, a special program (called
program “0”) is created, followed by a second program that contains
the run conditions for the actual run. Follow the instructions below.
ENTERING PROGRAM 0
The first program —program 0— is used to delay the start of the run
until the required time. Enter the settings as follows.
1. Press
_463+a
then
_a
on the keypad to access program 0. A
blinking zero appears above the
_463+a
key. The existing settings
for program 0 are displayed.
2. Use the parameter keys and keypad to enter time and temperature
settings as required. You cannot change the speed setting, which
is set at zero. (If you try to change it, the instrument will signal
an error.)
3. Press
. The SAVE message blinks to indicate that
you can save the program settings.
4. Press
_7%:)a
to save program 0.
5. To exit from the program mode, press
_')a
twice.
RUNNING PROGRAM 0 WITH A SECOND PROGRAM
To run the delayed start program function, do the following.
1. Press
_463+a
then
_a
on the keypad to access program 0. A
blinking zero appears above the
_463+a
key, and the display shows
the program 0 settings.
2. Use the keypad to select a second program from the stored
programs. The selected program settings appear in the display
and the program number appears in the
_463+a
display. Press
.
3. To begin the delayed start program function, press
and
_78%68a
.
The word “delay” appears in the SPEED display and the second
program number you selected appears in the
_463+a
display.