3-4
Operation
INSTALLING THE ROTOR
The power must be turned on before you can unlock and open the
chamber door. (To end a run for any reason, do not turn the power
switch off; press [
STOP
] instead.)
Action
Result
1. Turn the power switch on.
Indicator lights on the control panel light up.
2. Press [
DOOR
] to unlock
the door.
The instrument will accept this command only when the rotor is
at rest.
3. Lift the door up to open.
4. Install the rotor according
to directions in the rotor
manual. Ensure that the rotor
is seated on the drive hub.
Avoid bumping the control
panel keys during rotor
installation or removal.
(If you bump the control
panel and an L1 diagnostic
appears, press [
CE
] to
clear the message.)
!
CAUTION
Rotors used in Avanti J series centrifuges must
have drive pins in the rotor drive hole. These
drive pins engage with the centrifuge drive hub
to ensure that the rotor does not slip during
acceleration. Some Beckman Coulter rotors
were originally manufactured without drive pins
because these rotors did not need pins to run in
older model centrifuges. To check for drive pins,
hold the rotor up or turn it on its side and look
into the drive hole. If you do not see at least two
metal pins near the top of the hole, do not use
the rotor in the Avanti J-E.