Daily Start Up
3.5 Check Reagent Status
3
6
Select
Home > Reagent Management
.
7
Select
Reagent Check (F5)
. The Reagent Check window opens.
8
Select the unit(s) and check method to perform. Select
Start
to begin
the reagent check. Review
Main
and
Detail
to verify the results.
CAUTION
When the barcode label on a reagent bottle is dirty or has moisture
on it, barcode reading errors can occur. Check the barcode label of
the reagent bottle.
If the barcode label is dirty or has moisture adhering to it, wipe it off.
If barcode cannot be read even after wiping off dirt or moisture,
enter barcode ID manually. For details on ID editing, refer to
in
For precautions on filling bottles with reagent or Wash Solution, refer to 5.5.3
Reagent Replacement in the AU5800 User's Guide, chapter 5.